Error messages and trouble shooting
50
Troubleshooting
T
his section describes causes and actions to take for problems that customers frequently ask.
The indicated value fluctuates
< Problem with the electrode >
< Problem with the instrument >
< Problem with the sample >
The response is slow
Cause
How to solve problem
The conductivity electrode is dirty.
Clean the conductivity cell.
The conductivity electrode is broken.
Replace the conductivity cell.
There are air bubbles between the
conductivity electrode sleeve and
metal rings.
Shake the conductivity electrode to remove the air
bubbles.
Cause
How to solve problem
There is a motor or other device
causing electrical interference.
Measure at a place where no influence from
induction is given. Ground all AC-powered
equipment.
The conductivity electrode is not
connected correctly.
Connect the conductivity electrode properly.
Cause
How to solve problem
Some effects of the sample.
Response time may slow down, depending on
the properties of the sample solution.
Cause
How to solve problem
The conductivity electrode is dirty.
Clean the conductivity cell.
The conductivity electrode is broken.
Replace the conductivity cell.
Some effects of the sample.
Response time may slow down, depending on the
properties of the sample solution.