Getting Started

Installation Requirements

Locate the HotBlock® TKN under a fume hood with a minimum face velocity of 100 
fpm, and allow a minimum of 2” of space on all sides. The following environmental 
conditions should be observed: 

•  Ambient temperature range: 

5 to 30°C

•  Ambient relative humidity: 

0 to 90%RH

•  Altitude: 

sea level to 2500 meters 

HotBlock® TKN is rated as

 Pollution Degree 2

 and

 Installation Category 2. 

Electrical Requirements

Required Voltage:  

240 volts, ~60Hz, 15A

Power should not vary greater than ±10%. Use the supplied heat-resistant power 
cord or equivalent to connect to the power supply.
For safety reasons, a separate power receptacle should be provided for each unit in 
the system. Do not use extension cords or outlet adapters. Make certain that power 
outlets are earth-grounded at the grounding pin.

Содержание HotBlock TKN System

Страница 1: ...HotBlock TKN System Operation Instruction Manual 2021 Environmental Express Inc All rights reserved July 2021 1065EE1_M Rev 4 1 800 343 5319 or 1 843 881 6560 envexp com...

Страница 2: ...ing Your HotBlock TKN 6 Safe Sample Temperature Protection 6 Care Maintenance Maintenance 7 Troubleshooting 7 Circuitry Diagram 8 Replacement Parts Supplies 9 Product Information Item Date of Purchase...

Страница 3: ...isclaims all other warranties Environmental Express is not liable for any consequential special indirect or compensatory damages arising from use of or in conjunction with this product the maximum lia...

Страница 4: ...grounding pin See electrical requirements for power specifications page 4 Application of the wrong supply voltage can create a fire hazard and a potentially serious shock hazard and could seriously d...

Страница 5: ...ge alternating current Hz frequency A amperes Unpacking Your HotBlock TKN System 1 Remove the HotBlock TKN from the shipping container by lifting from the bottom of the block 2 If you purchased a comp...

Страница 6: ...e range 5 to 30 C Ambient relative humidity 0 to 90 RH Altitude sea level to 2500 meters HotBlock TKN is rated as Pollution Degree 2 and Installation Category 2 Electrical Requirements Required Voltag...

Страница 7: ...ion area Advanced Touch Screen Controller 1 Turn on the controller wait for the opening screen to appear 2 Touch the screen to proceed to the Main Menu Screen 3 Select the Global Settings menu 4 Touch...

Страница 8: ...does not have active cooling and cannot reliably use a time function to go down in temperature set point b Soak Use to hold the block at a temperature for a specific length of time c End Signifies the...

Страница 9: ...g can severely damage electronic components The controller is independent from the block and should be placed outside the fume hood to reduce the likelihood of acid damage Troubleshooting The HotBlock...

Страница 10: ...1 800 343 5319 or 1 843 881 6560 envexp com Environmental Express 8 HotBlock TKN Operation Instruction Manual Circuitry Diagram...

Страница 11: ...rts Supplies Description Item Advanced Touch Screen Controller TKN180 Simple Manual Controller TKN100 Replacement RTD Sensor TKN942 Replacement Heater Mat 54 Well Block TKN966 Replacement Heater Mat 2...

Отзывы: