background image

CL 200 - Portable Chlorine Colorimeter

Instruction Manual 

99561-78

Environmental Express

www.environmentalexpress.com

Содержание 99561-78

Страница 1: ...CL 200 Portable Chlorine Colorimeter Instruction Manual 99561 78 Environmental Express www environmentalexpress com...

Страница 2: ...onsible manner This service excluding the cost of transportation is free of charge This service only applies to electrical instruments purchased after 13th August 2005 Send your electrical Environment...

Страница 3: ...this can lead to incorrect test results 8 Contamination of the transparent cell chamber can result in wrong readings Check at regular intervals and if necessary clean the transparent cell chamber usi...

Страница 4: ...n 0 01 A Auto OFF automatic switch off 25 minutes after last keypress different reaction times depend ing on available method Display backlit LCD on keypress Time real time clock and date Calibration...

Страница 5: ...V1_04 2022 5 Guidelines for colorimetric measurements...

Страница 6: ...ent is waterproof seal ring E must be in position battery compartment cover B must be fixed with the four screws If the batteries are removed for more than one minute the date and time menu starts aut...

Страница 7: ...s whereby the zero adjustment is carried out in a 24 mm round vial with sample water It can be used if different tests are carried out with identical test conditions and the same water sample The zero...

Страница 8: ...CL 200 Instruction Manual 8...

Страница 9: ...le Cells 24 mm Glass 12 Pack with lids light shielding rings 99561 99 Required Materials Description Part Number Chlorine Reference Standard Kit 0 0 02 1 0 mg l Verification Standards 99561 98 Primary...

Страница 10: ...ld be free of chlorine consumption To achieve this all glassware should be placed in a sodium hypochlorite solution 0 1 g L for one hour and then rinsed thoroughly with deionized water 2 For individua...

Страница 11: ...24 mm vial with 10 ml sample Close vial Place sample vial in the sample chamber Pay attention to the positioning Press the ZERO button Remove the vial from the sample chamber Add Chlorine FREE DPD F10...

Страница 12: ...00 Instruction Manual 12 Place sample vial in the sample chamber Pay attention to the positioning Press the TEST button The result in mg l free chlorine appears on the display Chlorine PP 0 02 2 mg l...

Страница 13: ...24 mm vial with 10 ml sample Close vial Place sample vial in the sample chamber Pay attention to the positioning Press the ZERO button Remove the vial from the sample chamber Add Chlorine TOTAL DPD F1...

Страница 14: ...chamber Pay attention to the positioning Press the TEST button Once the reaction period is finished the measurement takes place automatically The result in mg l total Chlorine appears on the display...

Страница 15: ...ults within the measuring range of up to 0 mg L In this case the sample must be diluted with chlorine free water 10 ml of the diluted sample should be mixed with the reagent and the measurement taken...

Страница 16: ...CL 200 Instruction Manual 16...

Страница 17: ...mm Plastic 12 Pack with lids 99562 01 Required Materials Description Part Number Chlorine Reference Standard Kit 0 0 02 1 0 mg l Verification Standards 99561 97 Primary Chlorine Standard Kit EA makes...

Страница 18: ...d be free of chlorine consumption To achieve this all glassware should be placed in a sodium hypochlorite solution 0 1 g L for one hour and then rinsed thoroughly with deionized water 2 For individual...

Страница 19: ...lace sample vial in the sample chamber Pay attention to the positioning Press the ZERO button Remove the vial from the sample chamber Close vial Invert several times to mix the contents 20 sec Place s...

Страница 20: ...CL 200 Instruction Manual 20 Press the TEST button The result in mg l free chlorine appears on the display Chlorine PP HR 0 1 8 mg l CL8...

Страница 21: ...lace sample vial in the sample chamber Pay attention to the positioning Press the ZERO button Remove the vial from the sample chamber Close vial Invert several times to mix the contents 20 sec Place s...

Страница 22: ...l in the sample chamber Pay attention to the positioning Press the TEST button Once the reaction period is finished the measurement takes place automatically The result in mg l total Chlorine appears...

Страница 23: ...are eliminated by the addition of EDTA Concentrations above 8 mg L Chlorine in the event of using Powder Packs can lead to results within the measuring range of up to 0 mg L In this case the sample mu...

Страница 24: ...CL 200 Instruction Manual 24 Other Instrument Settings...

Страница 25: ...V1_04 2022 25...

Страница 26: ...CL 200 Instruction Manual 26...

Страница 27: ...V1_04 2022 27...

Страница 28: ...CL 200 Instruction Manual 28...

Страница 29: ...V1_04 2022 29...

Страница 30: ......

Страница 31: ......

Страница 32: ......

Страница 33: ......

Страница 34: ......

Страница 35: ......

Страница 36: ...Environmental Express www environmentalexpress com...

Отзывы: