V1_04.2022
15
Chemical Method
DPD
Interferences
Persistant Interferences
•
All oxidizing agents in the samples react like chlorine, which leads to higher results.
Removeable Interferences
•
Interference from Copper and Iron (III) are eliminated by the addition of EDTA.
•
Concentrations above 2 mg/L Chlorine, in the event of using Powder Packs, can lead-
to results within the measuring range of up to 0 mg/L. In this case, the sample must be
diluted with chlorine-free water. 10 ml of the diluted sample should be mixed with the
reagent and the measurement taken again (plausibility test).
Interference
from / [mg/L]
CrO
4
2-
0.01
MnO
2
0.01
Method Validation
Limit of Detection
0.01 mg/L
Limit of Quantification
0.03 mg/L
End of Measuring Range
2 mg/L
Sensitivity
1.68 mg/L / Abs
Confidence Intervall
0.033 mg/L
Standard Deviation
0.014 mg/L
Variation Coefficient
1.34 %
Conformity
EN ISO 7393-2
Chlorine PP / 0.02 - 2 mg/l / CL2