Environmental Express 35643-20 Скачать руководство пользователя страница 9

 

- 7 - 

necessary, calibrate it according to standard value measured by aneroid barometer. Refer to parameter 

setting P4.3 for the procedure of aneroid barometer calibration. 
5.2.5

 

Salinity compensation 

The instrument has manual salinity compensation. It is set in the parameter P4.2 (0 to 45 ppt). To obtain 

accurate  readings  in  mg/L  and  ppm,  it  is  necessary  to  know  the  salinity  of  the  test  sample  and  input 

salinity  value  into  the  instrument.  As  the  salinity  of  the  solution  increases,  the  level  of  DO  decreases. 

Generally, salinity of freshwater is 0 to 0.5ppt, salinity of seawater is 35ppt. 
 

5.3

 

DO Calibration 

 

5.3.1

 

Saturated oxygen calibration 

a)

 

This  procedure  requires  the  use  of  a  calibration  sleeve  to  allow  the  probe  to  be  calibrated  in  a 

humidity-saturated atmosphere. 

b)

 

Check  if  the  sponge  in  the  calibration  sleeve  is  damp.  Attach  the  calibration  sleeve  to  the  probe. 

Tighten the locking cap. Be careful not to have water droplets on the head of the sensor cap. Wait for 

5 to 10 minutes after turning on the instrument in order to saturate the air in the calibration sleeve 

with water vapor. In addition, wait for the temperature to completely stablize. 

c)

 

Long press        to enter the calibration mode, and 

CAL 

is flashing in the upper right corner. Wait for 

the stable icon        to appear and stay on LCD, press              to calibrate. Once

 100% 

becomes stable, 

the saturated oxygen calibration is completed. 

 

5.3.2

 

Zero-oxygen calibration   

Zero-Oxygen  calibration  is  only  performed  when  a  probe  or  sensor  cap  is  replaced,  the  probe  has  not 

been  used  for  a  long  time.  Generally,  it  is  not  necessary  to  perform  zero  oxygen  calibration.  The 

instrument  has  been  calibrated  for  zero  oxygen  before  it  leaves  the  factory,  so  it  is  not  necessary  to 

perform  zero-oxygen  calibration  for  the  first  use.  The  zero-oxygen  calibration  is  carried  out  in  the 

following steps. 

a)

 

Preparation  of  100ml  of  oxygen-free  water:  in  the  100ml  beaker,  weigh  2g  of  anhydrous  sodium 

sulfite (Na

2

SO

3

) and add 100ml of distilled or deionized water to dissolve. Oxygen-free water is only 

effective within 1 hour. 

b)

 

Put the electrode into the oxygen-free water, wait for 3 to 5 minutes after the instrument is turned 

on, and wait for the temperature and DO reading to completely stabilize. The DO reading should be 

very close to 0.1mg/L. 

c)

 

Long          press to enter the calibration mode. 

CAL

 will be flashing at the bottom right corner. Wait 

for a stable        . Press              to calibrate and the zero-oxygen calibration is completed.   

 
5.4

 

Special notes for calibration 

a)

 

Optical  dissolved  oxygen  probes  have  better  stability  and  smaller  calibration  drift  compared  to 

conventional electrochemical dissolved oxygen electrodes, which means that the meter can maintain 

Содержание 35643-20

Страница 1: ...DO1000 Portable Optical DO Meter 35643 20 35643 30 Kit Instruction Manual Environmental Express www environmentalexpress com...

Страница 2: ...5 2 Calibration preparation 5 3 DO calibration 5 4 Special notes for calibration 5 5 Measurement of water samples 5 6 Self diagnosis information 6 Parameter Setting 6 1 Main menu of parameter setting...

Страница 3: ...except for mechanical deterioration such as scratches to the light shielding layer the sensor cap has up to 8000 hours of service life Easy to calibrate and maintain Probe is equipped with a calibrat...

Страница 4: ...utomatic Salinity compensation range 0 45 ppt Manual Temperature Range 32 212 F 0 100 C Resolution 0 1 F 0 1 C Accuracy 0 9 F 0 5 C 1 digit Other Power AA batteries x3 1 5V 3 IP rating IP57 dustproof...

Страница 5: ...In measurement mode press to turn on off the backlight In calibration mode press to calibrate In the parameter setup mode press to enter and open the sub menu confirm parameter Short press In paramet...

Страница 6: ...s of nonuse Select Off to turn off automatic power off 3 4 2 Back light The meter is with white backlight suitable for application even in dark environments However when backlight is on the consumptio...

Страница 7: ...needs to be hydrated Soak the electrode in tap water at room temperature 77 F 25 C for 24 hours If the water temperature is low soaking time should be 48 72 hours Short term storage less than 30 days...

Страница 8: ...instrument The specific input method will be described in the instruction manual of sensor cap 4 3 6 The probe that comes with the meter can be used directly User should not take off the sensor cap wh...

Страница 9: ...rner Wait for the stable icon to appear and stay on LCD press to calibrate Once 100 becomes stable the saturated oxygen calibration is completed 5 3 2 Zero oxygen calibration Zero Oxygen calibration i...

Страница 10: ...re constant stirring of the solution or flowing fluid 5 5 2 Users can read the measurements when appears and stays on Note that the measurement time is related to temperature When the solution tempera...

Страница 11: ...linity value press to confirm 2 Press to enter P4 3 or press to return P4 3 Air pressure adjustment 60 120 kPa 1 Press 101 3 flashes press or to adjust air pressure value according to laboratory barom...

Страница 12: ...lashes press to select OFF On press to confirm Off Unlock up On lock up Stable reading 10 seconds and the reading will be automatically locked up 2 Press to enter P6 3 or press to return P6 3 Select b...

Страница 13: ...to finish 3 Check the sensor cap If it is contaminated clean it If it is dried out hydrate it If it is damaged replace it DOreadings are not accurate 1 Check whether the temperature is stable the sali...

Страница 14: ...ectrode cable is damaged or twisted due to external force c Fails to maintain the product in accordance with the requirements of this manual and the industry standard process d Any unauthorized repair...

Страница 15: ...96 8 45 51 8 11 0 11 03 10 42 9 84 9 29 8 77 8 28 53 6 12 0 10 78 10 18 9 62 9 09 8 59 8 11 55 4 13 0 10 54 9 96 9 42 8 90 8 41 7 95 57 2 14 0 10 31 9 75 9 22 8 72 8 24 7 79 59 0 15 0 10 08 9 54 9 03...

Страница 16: ...79 113 0 45 0 5 93 5 67 5 41 5 17 4 94 4 72 Salinity Dissolved salts in water Chlority Measure of chlorine content by mass of water S 1 80655 x Chlorinity Appendix 4DO Calibration Values Cal Value Air...

Страница 17: ...onmental Express 2345A Charleston Regional Parkway Charleston South Carolina 29492 Toll Free 800 343 5319 Technical Support 800 745 8218 Email tech environmentalexpress com Website www environmentalex...

Отзывы: