Environmental Express 35643-20 Скачать руководство пользователя страница 7

 

- 5 - 

 

 

1. Bottom cover of the calibration sleeve 

 

5. Locking cap 

 

2. Sponge for water storage 

 

6. Optical DO electrode 

 

3. Sensor cap 

 

7. O-ring 

 

4. Calibration sleeve 

 

Immersion line: The tested solution should be above this line   

 

4.2

 

Probe maintenance 

The sensor cap of the optical DO electrode must be kept in a moist environment. If the surface coating of 

the sensor cap dries out, the reading will be unstable, or the response rate will be slow.   

 

Before  the  first  use,  unscrew  the  bottom  cover  of  the  calibration  sleeve  to  check  if  the  sponge  is 

damp. If the sponge is dry or if the electrode is exposed in dry air for more than 8 hours, the surface 

coating of the sensor cap may be completely dried and needs to be hydrated. Soak the electrode in 

tap water at room temperature 77°F(25°C) for 24 hours. If the water temperature is low, soaking time 

should be 48-72 hours. 

 

Short-term  storage  (less than  30  days):  Keep the  probe  head in the  calibration  sleeve.  Always  keep 

the sponge inside the calibration sleeve moist. Several drops of clean water should be added to a dry 

sponge (let the sponge be damp, but not dipping), and tighten the lock cap, so that the sensor cap is 

kept in the moist-saturated air.   

 

Long-term  storage  (greater  than  30  days):  Keep  the  probe  head  in  the  calibration  sleeve.  Check 

whether  the  storage  sponge  is  moist  every  30  days  or  user  can  store  the  electrode  in  a  beaker 

containing clean water. 

 

The sponge cannot be allowed to get stained or moldy, otherwise it will consume or produce oxygen. 

If stained or moldy, please clean it or replace with new sponges immediately. 

 

Diagram- 9 

 

Содержание 35643-20

Страница 1: ...DO1000 Portable Optical DO Meter 35643 20 35643 30 Kit Instruction Manual Environmental Express www environmentalexpress com...

Страница 2: ...5 2 Calibration preparation 5 3 DO calibration 5 4 Special notes for calibration 5 5 Measurement of water samples 5 6 Self diagnosis information 6 Parameter Setting 6 1 Main menu of parameter setting...

Страница 3: ...except for mechanical deterioration such as scratches to the light shielding layer the sensor cap has up to 8000 hours of service life Easy to calibrate and maintain Probe is equipped with a calibrat...

Страница 4: ...utomatic Salinity compensation range 0 45 ppt Manual Temperature Range 32 212 F 0 100 C Resolution 0 1 F 0 1 C Accuracy 0 9 F 0 5 C 1 digit Other Power AA batteries x3 1 5V 3 IP rating IP57 dustproof...

Страница 5: ...In measurement mode press to turn on off the backlight In calibration mode press to calibrate In the parameter setup mode press to enter and open the sub menu confirm parameter Short press In paramet...

Страница 6: ...s of nonuse Select Off to turn off automatic power off 3 4 2 Back light The meter is with white backlight suitable for application even in dark environments However when backlight is on the consumptio...

Страница 7: ...needs to be hydrated Soak the electrode in tap water at room temperature 77 F 25 C for 24 hours If the water temperature is low soaking time should be 48 72 hours Short term storage less than 30 days...

Страница 8: ...instrument The specific input method will be described in the instruction manual of sensor cap 4 3 6 The probe that comes with the meter can be used directly User should not take off the sensor cap wh...

Страница 9: ...rner Wait for the stable icon to appear and stay on LCD press to calibrate Once 100 becomes stable the saturated oxygen calibration is completed 5 3 2 Zero oxygen calibration Zero Oxygen calibration i...

Страница 10: ...re constant stirring of the solution or flowing fluid 5 5 2 Users can read the measurements when appears and stays on Note that the measurement time is related to temperature When the solution tempera...

Страница 11: ...linity value press to confirm 2 Press to enter P4 3 or press to return P4 3 Air pressure adjustment 60 120 kPa 1 Press 101 3 flashes press or to adjust air pressure value according to laboratory barom...

Страница 12: ...lashes press to select OFF On press to confirm Off Unlock up On lock up Stable reading 10 seconds and the reading will be automatically locked up 2 Press to enter P6 3 or press to return P6 3 Select b...

Страница 13: ...to finish 3 Check the sensor cap If it is contaminated clean it If it is dried out hydrate it If it is damaged replace it DOreadings are not accurate 1 Check whether the temperature is stable the sali...

Страница 14: ...ectrode cable is damaged or twisted due to external force c Fails to maintain the product in accordance with the requirements of this manual and the industry standard process d Any unauthorized repair...

Страница 15: ...96 8 45 51 8 11 0 11 03 10 42 9 84 9 29 8 77 8 28 53 6 12 0 10 78 10 18 9 62 9 09 8 59 8 11 55 4 13 0 10 54 9 96 9 42 8 90 8 41 7 95 57 2 14 0 10 31 9 75 9 22 8 72 8 24 7 79 59 0 15 0 10 08 9 54 9 03...

Страница 16: ...79 113 0 45 0 5 93 5 67 5 41 5 17 4 94 4 72 Salinity Dissolved salts in water Chlority Measure of chlorine content by mass of water S 1 80655 x Chlorinity Appendix 4DO Calibration Values Cal Value Air...

Страница 17: ...onmental Express 2345A Charleston Regional Parkway Charleston South Carolina 29492 Toll Free 800 343 5319 Technical Support 800 745 8218 Email tech environmentalexpress com Website www environmentalex...

Отзывы: