Envirologix QuickSCan II Скачать руководство пользователя страница 17

13 

More about "Select Dilution": If you run a mycotoxin test kit with a 
high result, it can be further diluted, as per the Product Insert.  

For example, Aflatoxin 

 if no dilution was made, leave the selection 

at 1:1 (1 part in 1), or 
neat. The final result is 
11 ppb: 

If a dilution was made as per the Product Insert, click the down 
arrow in the Selection 
Dilution field.  Choose 
the dilution, as 
instructed, from the 
drop-down menu, in this 
case 1:6: 

Please note that in some cases, the dilution pulldown menu will 
display a letter instead of a number, for example "1:A" or "1:B". 

The program will then 
automatically calculate 
the result, 65 ppb: 

Please note, the results are given in 2 significant digits. In this 
example, while it could be assumed to be multiplying the 1:1 result 
of 11, it is actually multiplying an exact result that goes out several 
decimal places, for example, 10.8938. 

The Results Table features five user-input options, intended to enhance 
the traceability of deliveries or shipments. A drop-down menu for the 
user-entry columns will remember the last 30 entries made. Previous 
entries can be selected without re-typing them. The entries in the drop-
down menu can be removed by holding the Control key, arrow down to 
the item you wish to remove, and hitting the Delete key. Please note: if 
entries are defined in the Results Table Column Titles section of the 
Settings screen, they will re-appear in the next scan even if they are 
deleted here. If multiple test kits are scanned at one time, the program 
can autofill the entries for all kits after the first has been entered. You 
can override the autofill by re-entering the desired text in the 
appropriate box.  (See "Operation, Setup" for detail on Results Table 
Column Titles.) 

 

A scale-ticket number or trace identification number can be entered 
into the box directly under the "

Sample ID

" header 

 

A supplier name or number can be entered directly into the box 
under the "

Supplier

" header 

Содержание QuickSCan II

Страница 1: ...USER MANUAL ACC 331 Doc M247 0318...

Страница 2: ...ii EnviroLogix Inc 2018...

Страница 3: ...8 Inventory 9 Help 10 Exit 10 Reading Test Strips 10 Reading Multi Matrix Barcode Cards 11 Interpreting Results 12 Results Table 12 Save Report 14 Print Report 15 Print Receipt 15 Export 15 Help 15 Cl...

Страница 4: ...neous data storage where test results can be emailed printed or analyzed at any time A simple Microsoft Windows interface familiar to users worldwide and minimizing training time and A complete integr...

Страница 5: ...by genetic modifications in crops mycotoxins algal toxins chemical contaminants and plant pathogens Interest in detecting and measuring these targets lies in many industries including agricultural bi...

Страница 6: ...ion process contact the QuickScan Team at 1 866 408 4597 or techsupport envirologix com OPERATION Upon turning on the unit the QuickScan program will automatically open Or from the desktop double clic...

Страница 7: ...eport button See Print Report below for more details To print a Report highlight a row from the desired report and click the Print button A dialog window will ask if you wish to print DataLog rows or...

Страница 8: ...al output of your data depicting average results click the Analyze button Select test kit and starting and ending dates Select if you want running average and or supplier included on the graph Select...

Страница 9: ...older or network location Currently it defaults to C EnviroLogix QuickScan Export The Results Table Column Titles setting allows the user to edit the column headers of entry fields Select the original...

Страница 10: ...sers will see a daily reminder at 8 00 AM or upon startup thereafter to execute a Clean Test and Check Comb scan Changes to the time of the reminder can be made through Settings it is recommended that...

Страница 11: ...y ID will autofill the Location field with the Program Key ID If checked the Location field will not be editable QSS On Screen Keyboard select to activate the QSS keyboard when entering data into user...

Страница 12: ...s The cutoff will be applied to each individual result within the results screen even if multiple strips of the same test are scanned at one time Thus strips from multiple samples can be scanned at on...

Страница 13: ...ntory tracking then click the Add button The kit will appear in the Test Kit column of the Inventory grid Enter the number of tubes canisters for single strip kits or number of pouches for combs into...

Страница 14: ...by topic It also provides contact information for EnviroLogix Customer Service Each screen in the QuickScan program has a Help button When the Help button is clicked the program will open the file to...

Страница 15: ...ncoded as barcodes on the Barcode Card included in the kit box The curves must be uploaded into the program prior to running the test strips To do this open the drawer and slide the MMBC face up into...

Страница 16: ...he scale for mycotoxins is ppb or ppm For GMO tests results will fall between LOD the limit of detection and the upper limit of the quantitative range See the Product Insert for each test kit for more...

Страница 17: ...938 The Results Table features five user input options intended to enhance the traceability of deliveries or shipments A drop down menu for the user entry columns will remember the last 30 entries mad...

Страница 18: ...order in which strips are reported The QuickScan Results Screen provides a digital image of the scanned strip or comb The image shows the identification label barcode on the strip test lines and the s...

Страница 19: ...ck the Export button to create an export file Export file type can be selected from the Export Format drop down located in the Settings tab Export files can be exported to a location defined using the...

Страница 20: ...des with strip identifiers ID numbers unique to each device which are captured with the lot number within the DataLog When a product has been encoded with a Unique ID it is possible to clearly and qui...

Страница 21: ...ld not be removed from the strips and should not be shared or exchanged across different strips Altering the barcode in any way will affect performance of the QuickScan System When processing strips b...

Страница 22: ...date received if needed CLEANING AND SERVICING QuickScan II is designed to minimize dust or debris entering the scanner area and interfering with the reading of test strips It is important to keep th...

Страница 23: ...urface place the White Card into the drawer and click the Clean Test button on the main screen A record of the Clean Test will be stored in the DataLog Scan the Check Comb Place the Check Comb into th...

Страница 24: ...d in tray incorrectly Reposition strips and read again DIMENSIONS AND WEIGHT Length 37 cm Width 33 5 cm Height 20 cm Weight 11 15 lbs 5 06 kg SAFETY AND PRECAUTIONS 1 QuickScan is a precision system t...

Страница 25: ...n writing of such defects during the warranty period and returns the unit to EnviroLogix for evaluation EnviroLogix will repair or replace the unit at its option During the period of defects evaluatio...

Отзывы: