Identify, and confirm, the position of each of the component peaks on the chromatogram. The
retention time (RT) of each peak should be close (but does not need to be exact) to the value
listed in the calibration table on the Components page. If needed, make small adjustments to
the retention time of each peak’s location. The operator must ensure they have properly
identified each peak which eluted, and the order that they were detected
Prior to making any severe adjustments to the Retention Times, contact Envent Engineering
Ltd.
7.
Select one of the stable chromatograms and enter calibration mode by clicking the “Cal.”
Button on the toolbar at the bottom of the screen. An interactive system will open showing all
the timed events, right-click the chromatogram and click “Forced Cal.” Read and acknowledge
the dialogue box that pops up to allow the forced calibration to take place