Envent Engineering Ltd – 131S & 132S Software Manual – Revision 1
Page 47
3.
Automatic calibration is configured once written to the analyzer and will run on the interval
specified
4.
The calibration will automatically do 3 runs at the specified interval and generate a calibration
report when finished. If the calibration is a pass, the calibration factors will be updated, if it is
a failure, they will not be updated
5.
Calibration reports can be pulled from the analyzer as required
6.3.2
Manual Calibration
If the analyzer is not equipped with automatic calibration or a manual calibration is desired, follow
these steps.
1.
Navigate to the GC Control page under Operations
2.
Click the gray button for the calibration stream that needs to be run, this will queue the
calibration to be run next – do not click the Continuous check box, that is for forced
calibration
3.
The calibration automatically does 3 runs and generates a calibration report when finished. If
the calibration is a pass, the calibration factors will be updated, if it is a failure, they will not be
updated
4.
Navigate to the Components screen (Configuration > Components) and read from the device
(
) to ensure that the Response Factors update
5.
Calibration reports can be pulled from the analyzer as required