21
With the key [I] you can choose between different
information windows. The first window shows the raw
values, temperature and the status of the sensor.
The second window displays the error memory.
Recent error codes always come first. If an error code
is repeated, it will appear first, but will not be listed
multiple times.
If any data has been changed, the change will only be
taken into account when you exit the complete menu
structure and answer [ Yes ] when asked if you wish to
save the changes.
For reasons of simplicity, a further display menu screen has been dispensed with.
The display screens are directly derived from the menu structure in section 6.5.
Protection against unauthorised use:
If a password has been entered in menu
7. System
in
7.6 Password
, which is different to the “0000”
default setting, you will be asked to enter a password when attempting to access the menus.
After the password has been successfully entered, the menus will be unlocked for approx. 5 minutes
(from the last menu entry).
Save changes?
Y
N
Sensor Status
Temp Raw value
Stat
S1 --- 0.000123 OK
S2 --- 0.000213 OK
S3 --- 0.000321 OK
Average 0.000219