6.2 Basic information on the user interface
The operator interface differs depending on the system design:
•
DIN Rail housing without display, operation via PC software
•
Field housing with display, alternative operation via PC software
•
One to three sensor system
First of all, the different system versions are described below. Following that, the basic operation of the
ProSens system as a one sensor system is then described without going back over the different versions.
6.3 Differences between the DIN Rail and field housing Evaluation unit
The Evaluation unit in the DIN Rail housing is only a part of the functions available in the field housing.
The ProSens with converter only supplies one dust trend signal.
The following overview clarifies the differences between the versions.
Function
Field housing
DIN Rail
Converter
Menu system
•
via PC software
yes
yes
yes
•
via display
yes
no
no
Measurement value display current output
yes
yes
yes
Pulse output for measurement value output
yes
yes
no
Alarm system relay output
yes
yes
yes
Remote control digital input
yes
no
no
Cleaning pulse
yes
yes
no
Autocorrect analogue input
yes
no
no
Error output
•
on current output
yes
yes
yes
•
at relay
yes
yes
yes
•
via PC software
yes
yes
yes
•
via display
yes
no
no
•
at status LED
no
yes
yes
Measurement value output mg/m
3
or %
yes
yes
no
The Evaluation unit in the DIN Rail can only be configured via a USB or Modbus RS 485 interface and
PC program. On the Evaluation unit in the field housing, all functions can be configured by menu via
the touch-sensitive display. The field housing Evaluation unit can also be configured by PC.
The menu items on the display and in the PC software are numbered in a uniform manner so that
they can be referred to later on.