28
4.
Analog output
Settings and calibration of analogue output
4.1
Lower limit
Input:
0 … 22 mA
Standard: 3.2 mA
4.2
Upper limit
Input:
0 … 22 mA
Standard: 21 mA
4.3
Alarm value
Input:
0 … 22 mA
Standard: 2 mA
4.4
Alarm mode
Selection:
Alarm:
Alarm value at output, during alarm
alarm / freeze
measuring value is 0.
Freeze:
Last measurement value would be
freeze on Analogue output, till the error is
fixed.
4.5
Analogue output 1
Submenu
4.5.1 Calibration 4 mA
Selection:
The current output can be set via key
set output current
functions and adjusted at the receiving end.
4.5.2 Calibration 20 mA
Selection:
The current output can be set via key
set output current
functions and adjusted at the receiving end.
4.6
Analogue output 2
Submenu
4.6.1 Calibration 4 mA
Selection:
The current output can be set via key
set output current
functions and adjusted at the receiving end.
4.6.2 Calibration 20 mA
Selection:
The current output can be set via key
set output current
functions and adjusted at the receiving end.