MN-28929 (REV. 4) 06/15 • IVNOW MANUAL • 4
D A N G E R
Ensure power source matches
voltage identified on appliance rating
tag. The rating tag provides essential
technical information required for any
appliance installation, maintenance
or repairs. Do not remove, damage or
modify the rating tag.
Hazardous Voltage Present
Safety Class I
Equipment
Protective
Earth
Ground Symbol
Medical Equipment classified by Underwriters
Laboratories with Respect to Electrical Shock,
Fire and Mechanical Hazards only, in Accordance
with UL 60601-1 and CAN/CSA C22.2 No. 601.1.
UL File No.
E201645
Grounding reliability can only be achieved when equipment is
connected to an equivalent receptacle marked “Hospital Grade.”
ATTENTION
Consult accompanying
documents
D A N G E R
Do not
use this warming appliance
in the presence of flammable
anesthetic mixture (with air or with
oxygen or nitrous oxide).
This could cause an explosion!
(Not category AP or APG equipment )
ivNow-1
POWER REQUIREMENTS
120 V.A.C. — 60 Hz, 1 ph
0.15 kW, 1.3 Amps
Safety Class I Equipment
No Applied Parts
Mode of Operation: Continuous
NEMA 5-15P
15A - 125V Plug
Hospital Grade
ivNow-2
POWER REQUIREMENTS
120 V.A.C. — 60 Hz, 1 ph
0.3 kW, 2.5 Amps
Safety Class I Equipment
No Applied Parts
Mode of Operation: Continuous
NEMA 5-15P
15A - 125V Plug
Hospital Grade
ivNow-3
POWER REQUIREMENTS
120 V.A.C. — 60 Hz, 1 ph
0.45 kW, 3.8 Amps
Safety Class I Equipment
No Applied Parts
Mode of Operation: Continuous
NEMA 5-15P
15A - 125V Plug
Hospital Grade
ivNow-4
POWER REQUIREMENTS
120 V.A.C. — 60 Hz, 1 ph
0.6 kW, 5.0 Amps
Safety Class I Equipment
No Applied Parts
Mode of Operation: Continuous
NEMA 5-15P
15A - 125V Plug
Hospital Grade
ivNow-5
POWER REQUIREMENTS
120 V.A.C. — 60 Hz, 1 ph
0.75 kW, 6.3 Amps
Safety Class I Equipment
No Applied Parts
Mode of Operation: Continuous
NEMA 5-15P
15A - 125V Plug
Hospital Grade
ivNow-6
POWER REQUIREMENTS
120 V.A.C. — 60 Hz, 1 ph
0.9 kW, 7.5 Amps
Safety Class I Equipment
No Applied Parts
Mode of Operation: Continuous
NEMA 5-15P
15A - 125V Plug
Hospital Grade
E L E C T R I C A L I N F O R M A T I O N
The power specifications are located on the unit identification nameplate. This nameplate
is permanently attached to the unit and must be located to verify power requirements.
P R E P A R A T I O N
Before operating the module(s), clean the exterior of the unit with a damp cloth and general hospital
cleaner (isopropyl alcohol)
C A U T I O N
This unit has not been approved for
warming of blood or blood products.
WARNING
Injection Fluid manufacturer suggests not to
warm injection fluids ABOVE 40°C (104°F).
If fluids are warmed ABOVE suggested
temperature, they should be discarded.
Содержание ivNow Series
Страница 19: ...MN 28929 REV 4 06 15 IVNOW MANUAL 17 ivNow 1 120V Wiring Diagram ...
Страница 20: ...MN 28929 REV 4 06 15 IVNOW MANUAL 18 ivNow 2 120V Wiring Diagram ...
Страница 21: ...MN 28929 REV 4 06 15 IVNOW MANUAL 19 ivNow 3 120V Wiring Diagram ...
Страница 22: ...MN 28929 REV 4 06 15 IVNOW MANUAL 20 ivNow 4 120V Wiring Diagram ...
Страница 23: ...MN 28929 REV 4 06 15 IVNOW MANUAL 21 ivNow 5 120V Wiring Diagram ...
Страница 24: ...MN 28929 REV 4 06 15 IVNOW MANUAL 22 ivNow 6 120V Wiring Diagram ...
Страница 25: ...MN 28929 REV 4 06 15 IVNOW MANUAL 23 ivNow 1 230V Wiring Diagram ...
Страница 26: ...MN 28929 REV 4 06 15 IVNOW MANUAL 24 ivNow 2 230V Wiring Diagram ...
Страница 27: ...MN 28929 REV 4 06 15 IVNOW MANUAL 25 ivNow 3 230V Wiring Diagram ...
Страница 28: ...MN 28929 REV 4 06 15 IVNOW MANUAL 26 ivNow 4 230V Wiring Diagram ...
Страница 29: ...MN 28929 REV 4 06 15 IVNOW MANUAL 27 ivNow 5 230V Wiring Diagram ...
Страница 30: ...MN 28929 REV 4 06 15 IVNOW MANUAL 28 ivNow 6 230V Wiring Diagram ...