NT PRESSURE TRANSDUCER MODELS 4100, 4210
7
Entegris, Inc. | User Guide
PRESSURE LIMITS
—
TRANSDUCER
RANGE
MAXIMUM OVER PRESSURE
LIMIT @ 23°C (73°F)
MAXIMUM OVER PRESSURE
LIMIT @ 65°C (149°F)
0 – 30 psig
690 kPa (100 psig)
690 kPa (100 psig)
0 – 60 psig
1034 kPa (150 psig)
690 kPa (100 psig)
0 –100 psig
1034 kPa (150 psig)
690 kPa (100 psig)
0 –150 psig
1310 kPa (190 psig)
N/A
Maximum over pressure can be limited by the fitting. Consult the fitting specification for
maximum over pressure limits.
WARNING! The pressure limits
for standard NT Pressure Trans-
ducers (4000 Series) decrease
significantly for temperatures
above 65°C (149°F). Exceeding
these limits may result in personal
injury or equipment damage.
PRESSURE REFERENCE ACCURACY
—
The accuracy of the pressure
transducer output is ±1% of full
scale. This accuracy includes the
effects of linearity, hysteresis and
repeatability, measured at room
temperature. Accuracy specifica-
tions for non-standard product
configurations might vary.