Page 26 of 34
EN109-1631750-48 IFU
This unit operates with mains voltages. No attempt should be made to disassemble the unit.
Maintenance and repair should be carried out by authorized personnel only. The
manufacturer
will not be held responsible for the results of maintenance or repairs by unauthorized
persons.
On request a service manual can be made available containing: spare part list, descriptions, calibration
instructions and other information which will assist the user’s appropriately qualified technical
personnel to repair those parts of the equipment which are designated by the manufacturer as
repairable.
Restore Contact Control and Restore Factory Defaults
You can optimize the contact control yourself by means of the maintenance menu. Here you can also
restore the factory defaults.
Note: When restoring the factory defaults, previously stored values can be lost.
1)
Push System Settings
2)
Select “Maintenance” via
the
icon.
3)
Select the appropriate
item via the
icon.
Expected Service Lifetime
This device will remain suitable for its intended use as long as it is subjected to yearly maintenance by
a qualified service engineer as described in the service manual.
The service engineer will decide whether the device is suitable for use according to the specifications.
End of Life
This device contains materials which can be recycled and/or are noxious for the environment.
Please ensure that you are well informed of the local rules and regulations regarding
the removal of equipment and accessories.