background image

Enraf-Nonius B.V.

Radarmed 650 Service Manual

1409773

 

Description

January 2000

2-1

2

DESCRIPTION

Because of the modular construction of the Radarmed 650 the description is on block
diagram level.

2.1

BLOCK DIAGRAM (see appendix B fig. 1)

The electronic circuit of the Radarmed 650 is located on 5 modules:

-

Display module (also referred to as display PC Board).

-

Display interface module (also referred to as display interface PC Board).

-

Application module (also referred to as output stage PC Board).

-

Distributor module (also referred to as distributor PC Board).

-

H.F. output module (also referred to as magnetron). 

2.1.1

Display module (display PC Board)

The display module displays the H.F. output of the Radarmed 650 unit.  The signal used for
measuring the H.F. output is fed into the display PC Board via the display interface PC
Board.  The display PC Board is a three digit LED DC Voltmeter with an maximum resolution
of 1mV.  

2.1.2

Display interface module (display interface PC Board)

The display module adapts the anode voltage measured at resistors R125 / R126 on the
main PC Board.  Both resistors are parallel mounted in the anode circuit of the magnetron. 
The measured voltage is related to the current which flows through the magnetron and
therefore also related to the output power.  

2.1.3

Application module (main PC Board)

The function of the application module is to generate the output power to drive the
magnetron.

The application module comprises the following circuits:

-

Power supply circuits.

-

Fuses.

-

Output stages

Description of the application module:

The mains power is fed into the Radarmed 650 via a mains entrance.  Inside the mains
entrance two mains fuses are found.  The mains power runs via a mains filter and a mains
switch to the power supply PC Board.  As soon as the equipment is switched on the fan
starts to run and the magnetron is heated.  The mains supply is fed via the mains
configuration strip CNA (used for setting the magnetron heater voltage to 3.4V,  see par.
4.6.1) into the primary side of the mains transformer TR1.  The mains power is transformed
into 3.4V (magnetron heater), 8.8V, +/- 9.7V and fed into the power supply section of the
main PC Board.

 

For details and measuring points see appendix B figures 10, 11 and 12.

Содержание Radarmed 650

Страница 1: ...Radarmed 650 SERVICE MANUAL Part number 1409773...

Страница 2: ......

Страница 3: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Index January 2000 1 Radarmed 650 SERVICE MANUAL Copyright Enraf Nonius B V P O Box 810 2600 AV Delft The Netherlands part number 1409 773 40...

Страница 4: ...2000 3 1 3 2 TOOLS AND MATERIALS Jan 2000 3 1 4 MAINTENANCE Jan 2000 4 1 4 1 GENERAL Jan 2000 4 1 4 2 CHECKLIST Jan 2000 4 1 4 3 FUNCTIONAL TEST Jan 2000 4 1 4 4 TROUBLE SHOOTING Jan 2000 4 2 4 5 REP...

Страница 5: ...Chassis side view Jan 2000 B 4 Fig 5 Top cover inside Jan 2000 B 5 Fig 6 Main PC Board Jan 2000 B 7 Fig 7 Distributor PC Board Jan 2000 B 9 Fig 8 Display PC Board Jan 2000 B 11 Fig 9 Display interfac...

Страница 6: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Index January 2000 4...

Страница 7: ...oints out a statement deserving more emphasis than the general text but which does not deserve a Warning or Caution Warning Inside the Radarmed 650 high voltages are available beware of electrical sho...

Страница 8: ...batch numbers of main sub assemblies and or important parts per unit The parts listed in a configuration file are defined as registered parts Such are for example parts as transformers PC Boards key...

Страница 9: ...ons to report any structural problem which may be discovered When the same problem regularly occurs or when safety problems occur we would like you to send us a product complaint report The report mus...

Страница 10: ...mitter is used to treat large areas The special form of this emitter is well adapted to the appropriate body contours Restricted regions such as the shoulder and knee joints can likewise be advantageo...

Страница 11: ...t be connected to a wall socket with protective earth connection 2 B Equipment providing a particular degree of protection against electric shock according to IEC 601 1 and IEC 601 2 6 safety standard...

Страница 12: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Introduction January 2000 1 6...

Страница 13: ...ge measured at resistors R125 R126 on the main PC Board Both resistors are parallel mounted in the anode circuit of the magnetron The measured voltage is related to the current which flows through the...

Страница 14: ...ied to the distributor module can be switched by different combinations of connecting the wire Connecting the wire combinations will result in the following voltages CN3 to CN2 1900VAC CN6 2000VAC CN3...

Страница 15: ...high voltage at transformer TR2 The higher the voltage at the secondary side of TR2 the higher the voltage on the capacitor C1 and at the magnetron The current which flows through the magnetron is in...

Страница 16: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Description January 2000 2 4...

Страница 17: ...Rigel Safety tester model 233 3 2 TOOLS AND MATERIALS It is assumed that a complete set of precision engineering tools is available For the service of the Radarmed 650 the following additional tools a...

Страница 18: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Test equipment and tools January 2000 3 2...

Страница 19: ...ir is blown threw the magnetron and that the air intake is free Check that the three segment LED output power display will light display shows 000 Switch off the unit 4 3 2 Treatment timer Connect a l...

Страница 20: ...tests Switch on the Radarmed 650 unit Check the supply voltages connect a multimeter with the negative testing lead to ST168 ground terminal at the main PC Board see appendix B fig 6 Connect the posi...

Страница 21: ...autions should be observed Persons should be earthed by means of a wrist strap Ground all electrical equipment work bench soldering iron etc After every repair a functional test must be performed see...

Страница 22: ...emove the display PC Board Installation Installation of the display PC Board is in reverse order of the removal procedure 4 5 6 Output power control potentiometer R1 Removal See par 4 5 2 for removal...

Страница 23: ...front cover Remove the high voltage wiring from the distributor PC Board Remove the primary wiring from connector ST171 main PC Board and from resistor R2 Remove the four mounting screws Caution Be ca...

Страница 24: ...o ground is present during the anode current adjustment The following power adjustment refers to the use of a moving coil milliammeter in the DC range and a large 150 W emitter for the 100 W adjustmen...

Страница 25: ...est value is found the current can be fine adjusted by means of adjusting resistor R2 Fine adjustment is done by moving the slide over the wire wound resistor Before moving the slide loosen the screw...

Страница 26: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Maintenance January 2000 4 8...

Страница 27: ...must be stored at room temperature 20 C The relative humidity should be less than 50 The storage room must be clean dust free and ventilated For the spare parts of the Radarmed 650 the above mentioned...

Страница 28: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Storage and transport January 2000 5 2...

Страница 29: ...g data 6 1 1 Normal delivery The procedure for normal delivery is the standard procedure for processing your spare parts orders This procedure is suitable for spare parts orders which do not require r...

Страница 30: ...rvice General Information guide The procedure is as follows 1 Fill in the Authorization Request for Return Shipment form and send this form by fax to your agent at Enraf Nonius 2 Within two working da...

Страница 31: ...A number in this column indicates the minimum recommended quantity for spare parts stock at the local service organization Quantity of equipment in use 1 50 C A Y yes in this column marks the so call...

Страница 32: ...on F103 10 10 2655170 Fuse T 0 315A 250V 5X20mm F104 10 10 2655172 Fuse T 0 5A 250V 5X20mm F105 10 10 2655168 Fuse T 0 2A 250V 5X20mm F106 10 10 2655175 Fuse T 1A 250V 5X20mm FL 0 0 2670079 Filter mai...

Страница 33: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Appendix B January 2000 B 1 APPENDIX B DRAWINGS Figure 1 Block diagram...

Страница 34: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Appendix B January 2000 B 2 Figure 2 Radarmed 650 top view...

Страница 35: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Appendix B January 2000 B 3 Figure 3 Chassis front view...

Страница 36: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Appendix B January 2000 B 4 Figure 4 Chassis side view...

Страница 37: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Appendix B January 2000 B 5 Figure 5 Top cover inside...

Страница 38: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Appendix B January 2000 B 6...

Страница 39: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Appendix B January 2000 B 7 Figure 6 Main PC Board...

Страница 40: ...layer Product code s 3412319 Part no of complete assy printed on label on PC Board N A Part no of PC Board only printed on PC Board 0409809 Equivalent service spare part part no not printed on PC Boar...

Страница 41: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Appendix B January 2000 B 9 Figure 7 Distributor PC Board...

Страница 42: ...ulti bi layer Product code s 0409680 Part no of complete assy printed on label on PC Board N A Part no of PC Board only printed on PC Board 0409814 Equivalent service spare part part no not printed on...

Страница 43: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Appendix B January 2000 B 11 Figure 8 Display PC Board...

Страница 44: ...bi multi bi layer Product code s 2611120 Part no of complete assy printed on label on PC Board N A Part no of PC Board only printed on PC Board 2611120 Equivalent service spare part part no not printe...

Страница 45: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Appendix B January 2000 B 13 Figure 9 Display interface PC Board...

Страница 46: ...single bi multi single layer Product code s 0409681 Part no of complete assy printed on label on PC Board N A Part no of PC Board only printed on PC Board 0409681 Equivalent service spare part part n...

Страница 47: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Appendix B January 2000 B 15 Figure 10 Wiring diagram...

Страница 48: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Appendix B January 2000 B 16 Figure 11 Mains distributor strip configuration Figure 12 Power supply diagram...

Страница 49: ...as Service Info s product change notes Once every year Enraf Nonius publishes a Service Info which lists all the published Service Info s until sofar Once every year Enraf Nonius publishes a Service...

Страница 50: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Appendix C January 2000 C 2 C 2 Revision history First edition January 2000 Revisions Chapter Revision Date of issue page...

Страница 51: ......

Страница 52: ...Subsidiary of Henley Healthcare Inc Enraf Nonius B V P O Box 810 2600 AV Delft The Netherlands...

Отзывы: