Enraf Nonius MANUMAX Скачать руководство пользователя страница 17

 

Page 17 of 20 

EN109-1570750-41 IFU

 

9

 

Specifications 

 

Electrical Info  

Mains voltage 

100 - 240 Volt  

Frequency 

50/60 Hz 

Power consumption 

250 VA max 

 

 

Safety and performance standards 

 

Safety class 

II according to IEC 60601-1 

Moisture resistance 

IPX1 

Safe working load 

250 kg 

Medical classification 

Class I in accordance with MDD (93/42/EEC) 

 

 

Performance

 

 

Lifting capacity 

250 kg 

Height adjustment speed 

Approx. 2.5 cm/second 

 

 

Dimensions

 

 

 

Width   

Depending on the configuration, 67 or 80 cm 

 

Length 

Depending on the configuration, approximately 
205-215 cm 

 

Height   

Adjustable from approximately 48 to 96 cm 

 

Weight 

Depending on the configuration, approximately 
100 kg 

 

 

Upholstery 

Biocompatible 

In accordance with 10993-1 

Latex-free 

Yes 

Fire retardant 

DIN EN 1021 (optional Crib. 5 BS available) 

 
Subject to technical modifications 

 
Environmental conditions 

Environmental conditions for transport and storage 

Environment temperature:  

10 to +50° C 

Relative humidity:  

20 to 90% (non-condensing) 

Atmospheric pressure:  

700 to 1060 hPa 

Environmental conditions for normal use 

Environment temperature: 

+5 to +40° C 

Relative humidity: 

20 to 90 % (non-condensing) 

Atmospheric pressure: 

700 to 1060 hPa 

 

Содержание MANUMAX

Страница 1: ...Page 1 of 20 EN109 1570750 41 IFU MANUMAX EN109 1570750 41 IFU July 17 2019 Instructions For Use...

Страница 2: ...IFU Table of Contents 1 Introduction 3 2 Symbols 4 3 Package Contents 5 4 Installation 6 5 Intended use and intended user 8 6 Precautions 9 7 Operation 12 8 Maintenance 15 9 Specifications 17 10 Conta...

Страница 3: ...y of life of many people This Instructions For Use relates to the ManuMax Examination 2 2 the ManuMax Multi 3 3 and the ManuMax Multi 3 5 It is important to read the following instructions carefully b...

Страница 4: ...e safely exposed Waste electrical items that can be recycled Indicates the electrical and electronic components of the device can be recycled and has to be disposed separately Manufacturer name addres...

Страница 5: ...following is included with your treatment couch 1x 1570750 ManuMax User Manual PDF on CD 1x 1570751 ManuMax Information Booklet The use of accessories and cables other than those specified or supplie...

Страница 6: ...on The couch weighs approximately 100 kg and must be unpacked by at least two people Do not lift the couch by the table top parts or the all around control Immediately after unpacking the couch check...

Страница 7: ...he frame Hooks are fitted on the inside of the subframe to guide the power cord to one of the four legs Guide the cord to the most suitable leg depending on where the wall socket in your treatment roo...

Страница 8: ...ded to support a patient in the desired position for examination or treatment within the physiotherapy rehabilitation and adjacent disciplines Intended User The ManuMax treatment couches are intended...

Страница 9: ...ver thresholds The ManuMax treatment couches are intended to be used and shall only be used by or under the supervision of professional users in physiotherapy rehabilitation and adjacent areas Never a...

Страница 10: ...couch is 250 kg This means that the total weight on the couch patient weight plus any other load such as manipulation by the therapist must not exceed 250 kg Never leave children alone with this devi...

Страница 11: ...re not to make contact with the motor when placing the lift or emitter Placing objects under the couch with a height exceeding 95 mm may result in dangerous situations and serious damage Try to limit...

Страница 12: ...o select the desired mode One or more beeps will sound The number of beeps represents the selected setting Press once 1 beep sounds lock enabled time of 30 minutes Press twice 2 beep sounds lock enabl...

Страница 13: ...ax couches are supported by gas springs These ensure an easy adjustment and stable positioning of the table top sections However the gas springs are not strong enough to adjust a section while support...

Страница 14: ...nlock the leg section lever 3 and put the leg section in the angled position using the other hand Never adjust a table top section when the patient is supporting on it with his her full weight The tab...

Страница 15: ...ons as provided in the instructions for use of the disinfectant Never use a mix of disinfectants or a mix of disinfectants and soap Preferably apply the product to the imitation leather using a cloth...

Страница 16: ...ny mechanical damage to the couch do not use the couch and contact your supplier Expected Service Life Provided the annual maintenance is carried out by a qualified Enraf Nonius service technician or...

Страница 17: ...Length Depending on the configuration approximately 205 215 cm Height Adjustable from approximately 48 to 96 cm Weight Depending on the configuration approximately 100 kg Upholstery Biocompatible In a...

Страница 18: ...cluding without limitation damages for loss of goodwill work or productivity computer malfunctions and any other form of business damage or loss even if advised of the possibility of their occurrence...

Страница 19: ...Page 19 of 20 EN109 1570750 41 IFU...

Страница 20: ...EN109 1570750 41 IFU Copyright Enraf Nonius B V Vareseweg 127 3047 AT Rotterdam The Netherlands Tel 31 0 10 20 30 600 info enraf nonius nl...

Отзывы: