Page 39 of 46
EN109-1633750-40 IFU
Error Messages
When the device is turned on, it will first execute a self-test. When an error is detected, both during
the self-test and during normal operation, a pop-up screen will appear on the display. When the
error is displayed, all outputs will be disabled. When this situation occurs, switch off the unit and
disconnect all accessories. Wait approximately 15 seconds and switch the unit back on.
When the error re-appears, stop using the device and contact the supplier.
Device needs to cool down before you can proceed
The temperature of the device is too high to start treatment. Remove the device from any potential
heat sources (such as direct sunlight) and let the device cool down. Restart the device to continue.
Laser Probe Error
Laser Probe must be connected to the Laser Probe holder. Do not connect the Laser Probe directly
to the Endolaser 120.
Remote Interlock or Emergency stop activated
Restart the device by switching the power off and on. Check Remote interlock connections if
problem persists.
Unsupported Device Detected
Disconnect to continue.
Troubleshooting
The device should not be used when mechanical damage is noticed. This includes damaged
cable insulation.
In the event that an error message or warning appears, immediately stop all use of the device.
Shutdown and restart the device. If the problem persists contact the supplier or Enraf-Nonius
B.V. for service. Errors and warnings indicate an internal problem with the device that must be
tested by a field service technician certified by Enraf-Nonius B.V before any further operation
or use of the system.
Use of a device that indicates an error or warning may pose a risk of injury to the patient,
user, or extensive internal damage to the system.
In case of failure of the display or other obvious defects, unplug the device immediately and
notify a certified service technician.
In the case of ingress of liquids, unplug the device from the Mains supply and have it checked
by an authorized person.