Page 21 of 46
EN109-1633750-40 IFU
Operation
The Endolaser 120 has two independent output channels. On each channel you can set parameters.
Only one channel can be active at a time. You cannot work simultaneously with Laser Probe A and
B. The device will automatically recognize the connected laser probe type.
After setting time and dosage, the laser system switches from
laser stand-by
to
laser ready
. When the
laser system is in
laser ready
mode, the laser indicator (LED on the probe) turns green and the red
target light is visible. Pushing the trigger button on the Laser Probe will activate the treatment. The
laser indicator (LED on the probe) turns orange, indicating that emission is in progress during therapy
application. Pressing the trigger button again during treatment will pause the laser output. This state
will be shown by the pause icon of the run status indicator above the Time. Pressing the laser trigger
again will continue treatment.
When the laser is active, pushing the laser stop icon in the user interface or the emergency laser stop
button on the Laser Probe holder will terminate laser emission. To restart a treatment, parameters will
need to be set up again.
Navigation
The Endolaser 120 is equipped with touch screen technology. Select a menu item by touching the
button to navigate to the next screen. You can navigate back to the previous screen by touching
the back button at the top of the screen.
Home Menu
The Home menu gives access to all functions of the unit.
•
Select “Clinical Protocols” to start working with preset
protocols
•
Select “Favorites” to use a manually saved protocol
•
Select “Manual Operation” to set all parameters manually
•
Select “System Settings” to enter the settings menu
If no Laser Probe is connected only System Settings can be
selected.