5

 

 

  from the surface of specimen. Select the mode as Fig.4-2. 

 
 

4.1.2  Limits 
of tolerance 

As the limits 
of tolerance 
are set, once 
the 
measured 
value 
exceeds 
these limits 
the gauge 
will sound 
alarm. 

 
In b.Tolerance: 
Can set to On (turn limit alarm on) or 
Off(turn off limit alarm); 
Or can select    Upper Limit    or  Lower 
Limit, to input the value. 
 

Note

The upper limit should be greater 

than the lower limit, otherwise this 
alarm function is invalid. 
 
 
 
 
 

 

Fig4-3 

 

Fig.4-2 

Содержание ETG15F1

Страница 1: ...ETG15F1 Coating Thickness Gauge User s Manual POLYGON INSTRUMENT LIMITED ...

Страница 2: ...Browse 7 4 2 2 Print 7 4 2 3 Delete 8 4 2 4 Upload 8 4 3 System 9 4 3 1 Auto printing 9 4 3 2 Backlight 9 4 3 3 Beep 10 4 3 4 Power off 10 4 3 5 Time Date 10 4 3 6 Unit 10 4 3 7 Default curve 11 4 3 8 Default system 11 5 Maintenance 12 5 1 Calibration 12 5 1 1 Zero calibration 12 5 1 2 One point 13 5 1 3 Two points 14 5 2 Influence factors 15 5 3 Maintenance 16 Appendix A 1 Packing List 17 A 2 Spe...

Страница 3: ...lectroplating ship and bridge building aircraft construction and the engineering and chemical industry 1 1 Principle The coating thickness gauge work on the magnetic induction principle for measuring the thickness of non magnetic coatings such as paint enamel rubber aluminium chrome cupper tin etc on magnetic metal substrates ferrous bases and steel also on alloyed steel or on hardened magnetic st...

Страница 4: ...ance are shown as Fig 2 1 and Fig 2 2 2 1 LCD display Fig 2 1 Built in Sensor LCD USB Power Button Key Up Down Zero ESC Menu Enter Measured number Amount of test group Mean value Measured value Calibration curve number Measure mode Single Mode Continue Mode Fig 2 2 ...

Страница 5: ...Error Surroud Error Should take gauge far from magnetic metals and wait few seconds it can go normal 3 2 Measuring Procedure a Ready the test piece b Switch on the gauge 0 0μm mil is displayed c Calibrate the gauge if necessary d Measurement Press gauge on the measured surface lightly and vertically Or place the gauge on measured surface of specimen directly After the buzzer can lift the gauge and...

Страница 6: ...r return to the previous menu You can also press to exit setup and return to the measurement mode In value input Press to move digit press to increase or decrease value 4 1 Measuring 4 1 1 Measuring mode 1 Single In single mode the sensor of gauge touch the surface of specimen measure one thickness value For measuring the another point must lift up the gauge from surface of specimen no less than 1...

Страница 7: ...ce the measured value exceeds these limits the gauge will sound alarm In b Tolerance Can set to On turn limit alarm on or Off turn off limit alarm Or can select Upper Limit or Lower Limit to input the value Note The upper limit should be greater than the lower limit otherwise this alarm function is invalid Fig4 3 Fig 4 2 ...

Страница 8: ...sure precision of measurements if selected Form Data Press the certificated time and date will be displayed if a curve which not certificated is selected will display 4 1 4 Amount of the test groups The gauge saves the measured data in test groups the statistical evaluation is in one group too Each group can be 1 to 9 data Fig 4 5 Fig 4 4 Fig 4 5 ...

Страница 9: ...an browse or print them Select data then presss can disply some information of selected group 4 2 2 Print The gauge can connect to wireless printer for printing data in memory First of all turn on the printer and make sure nearer than 3m arround the gauge Fig 3 7 Fig 4 6 Calibrating Curve Mean value Standard Deviation Amount of measurements Group number Fig 4 7 ...

Страница 10: ...function you can delete current measured value current measured group or all data in memory NOTE It couldn t restore again once you deleted the data 4 2 4 Upload The data in memory can be uploaded to computer you can use the EDLS software to receive data And manage transform or analyse these data with computer EDLS software must be set up in your computer See manual of EDLS Fig 4 8 Fig 4 9 ...

Страница 11: ...d group and All 4 3 System 4 3 1 Auto printing If Auto printing is turned on print measured value of a group as the measuring end automatically When the gauge is restarted Auto Printing is defaulted to OFF 4 3 2 Backlight You can set Backlight to Off 15s 30s and On Set to Off shut the backlight down Set to On the backlight will be lit Fig 4 10 Fig 4 10 Fig 4 11 ...

Страница 12: ...tion Note Set the backlight to On isn t recommended for battery power save 4 3 3 Beep Fig 4 12 4 3 4 Power off If the auto power off is set to On power will be shutted down at 3 minutes after no any operation 4 3 5 Time Date Fig 4 14 4 3 6 Unit Fig 4 15 Fig 4 12 Fig 4 13 Fig 4 14 Fig 4 15 ...

Страница 13: ...he zero calibration and one point calibration and two points calibration are all deleted So make sure you want to set to the default curve 4 3 8 Default system Perform this function all factory settings are restored Calibration curves measured value in memory and settings by custom are all deleted So be careful to use this function Fig 4 16 Fig 4 17 ...

Страница 14: ...to bring a certain error appropriate calibration can improve its accuracy of measurement The gauge is designed with three levels of calibration methods zero calibration one point calibration and two point calibration 5 1 1 Zero calibration There are two ways for zero calibration measuring calibration and menus calibration 1 Measuring zero calibration Fig 5 1 Fig 5 2 ...

Страница 15: ...curacy 2 Menu zero calibration a Enter 1 Mesure e Calibration select Zero press to display Acting b Measuring on substrate directly 4 or 5 times measurements should be done for more accuracy C Press for save zero calibration or press to discard calibration Fig 5 2 5 1 2 One point calibration Before performing the one point calibration zero point must be calibrated a Enter e Calibration select 1st ...

Страница 16: ...point calibration you can do the two points calibration Select 2nd the following process is exactly the same as one point calibration Fig 5 4 Note The standard foil used in two points calibration should be thicker than the one used in one point calibration 5 2 Influence factors 1 Magnetic properties of the substrate The gauge used magnetic induction method is affected by variations in the magnetic...

Страница 17: ...e to make a greater number of measurements at different positions to obtain an average value that is representative of the mean coating thickness 4 Residual magnetism Residual magnetism in the basis metal affects the measurements made by gauge that employ a stationary magnetic field Its influence on measurements made by gauge employing an alternating magnetic field is much smaller 5 Stray Magnetic...

Страница 18: ...removed 6 When the batteries are low will be displayed The batteries should be changed immediately Note 3 1 5V alkaline batteries The calibrations and all kinds of settings are stored in a nonvolatile memory they will not be lost even if batteries go dead completely or during the process of changing the batteries If the gauge is not to be used for an extended period of time remove batteries to avo...

Страница 19: ...SB Cable 1 Operation manual 1 A 2 Specifications Test Method Magnetic induction Measuring Range 0 1500mm 0 60mil Resolution 0 999μm 0 1μm 1000μm 1μm Accuracy up to100μm 1 5μm 100 1500μm 1 5 Data Memory up to 1500 readings Display Graphic LCD with backlight Power Supply 1 5V 3 AAA batteries Environment 0 50 C Dimension 88 X 67 X 30 mm Weight 120 gr ...

Отзывы: