Installation and Operation Manual - SILOTEC 8000
Software Overview and Explanations
Doc.-ID: SILOTEC_11022020_EN
31
The ACAL can be started by:
Time:
Time based start with set intervals in days and the corresponding time
Time + Digital Input:
Same as "Time", additionally a control voltage of 12-24V DC must be applied to the "calibration
release" so that an automatic calibration can be started.
Digital Input:
A control voltage of 12-24V DC must be applied to the "calibration release" so that an automatic
calibration can be started. If the control voltage is still present after the calibration has ended, the calibration will
immediately restart.
Info
In order to obtain the most accurate possible calibration of the sensors, they should be calibrated
separately. A binary cylinder gas such as 2.1 % O
2
, Balance N
2
, is recommended for the O
2
sensor. The
test gas recommendation for the CO
e
sensor can be found in the enclosed test protocol.
ENOTEC REMOTE
4.2.12
Info
ENOTEC REMOTE is disabled by default. Password and range are not displayed. After activation (only
possible from system access level), the 8-digit password that is requested during the remote connection
setup must be assigned.
The ENOTEC REMOTE password is used for:
authentication and pairing with a smartphone / tablet / laptop / pc.
authentication / login after every connection. Without Authentication / login, device data cannot be read or
modified. The device configuration can also not be altered.
Range
limits the transmission power of the ENOTEC REMOTE module.
Maximum
= 100m,
medium
= 10m,
short
=
1m. The actual possible range may vary due to structural factors and the reception strength of the
Smartphone/Tablet.
When an ENOTEC REMOTE connection to the
analyzer is active, the connection is shown in the
upper right corner of the display
.
Figure 26
- ENOTEC REMOTE connection
active
Info
A maximum of 16 users (smartphones /tablets) can connect to the ENOTEC REMOTE
module of an
ENOTEC analyzer.
Should additional users attempt a connection, the connection will fail. In this case, manually switch off the
ENOTEC REMOTE and switch it back on again
SYS MENU => System configuration => ENOTEC
REMOTE Settings
which will reset the module.
All previously paired users will need to delete their saved connection to the analyzer and re-pair their
devices.
Measuring units
4.2.13
Measuring units can be set for temperature in °C and °F, for pressure in mbar and psi and for concentration in ppm or
mg/m³