RTU mode
address 011
Information in the Well Watch is accessed by register number using Modbus function 3
to read registers, 6 to write registers, or 16 to write multiple registers. It should be
noted that well watch registers are 32 bits wide and may represent integers or floating
point data. By default, Modbus assumes registers are 16 bits wide, and transmitted in a
character stream with bits 8-15 in the first character and bits 0-7 in the second. To
transfer a 32 bit register, the Well Watch assumes two 16 bit registers are transmitted
with the low order 16 bits in the first register then the high order.
The addresses are listed below.
All of the Well Watch data is sent as 32 bit words as either floating point
representations or as integers. The master device needs to be set accordingly.
Address space REGISTERS (read only):
100
Depth running ave
32 bit Float
101
Depth instantaneous
32 bit Float
102
Depth at power max
32 bit Float
106
Water in Well
32 bit Float
107
Depth last good
32 bit Float
110
Battery Voltage
32 bit Float
111
Probe Temp
32 bit Float
112
Error Code
32 bit Integer
113
Gain Pulse
32 bit Integer
114
Gain Echo
32 bit Integer
2000
Software version
32 bit Integer
2001
Build Number
32 bit Integer
2002
Configuration
32 bit Integer
2003
Features
32 bit Integer
2004
Serial Number
32 bit Integer
Address space REGISTERS (read/write):
1010
RS232 packed parameters
32 bit structure
1011
RS232 mode 32 bit
Integer
1012
RS232 baud
32 bit Integer
1013
RS232 data bits
32 bit Integer
1014
RS232 stop bits
32 bit Integer
1015
RS232 parity enable
32 bit Integer
1016
RS232 parity
32 bit Integer
Eno Scientific Well Watch 670
Page 26