ProAir 2200
ENMET
Manual Revision Date – May 3, 2017
P a g e
|
6
Manual Part No. – 80002-046
Figure 1: External Features of the ProAir 2200
N
OTE
: Typical gas reading & alarm locations, depending on instrument configuration, alarms & readings may be in alternate
locations.
Sample Air Hose, 5ft Supplied with Monitor
Example of Regulator,
Not Supplied with Monitor
Figure 1A: Regulator and Sample Air Hose
N
OTE
:
Typical facility airline pressures range from about 55 to 125
PSI
.
The outlet pressure of the regulator supplied with the monitor,
which is connected to the inlet of P
RO
A
IR
2200 the, should be
set at 55
PSI
,
±
5
PSI
.
Female Quick Release
to Inlet Port to P
RO
A
IR
2200
¼” NPT
N
OTE
:
When connecting to a standard 55
PSI
USA Medical air system, Regulator is Not required
.
Female Quick Release
to Sample Port
Male¼” NPT Fitting
To Air System
Male Quick Release Fitting
Input Port
Sample/Calibration
See note below
O
2
%
CO
ppm
Front Cover
Latch
Audio Alarm
Fault LED
SELECT
Switch
Power LED
OPTION
Switch
A
LARM
A
CKNOWLEDGE
/
A
LARM
D
EFEAT
Switch
*Visual Alarm O
2
(If Oxygen option is
installed)
*Visual Alarm CO
(See *note below)
Mounting Flanges
2 places
Dew
Point
*Visual Alarm Dew Point
(
If Dew Point option
is installed)
See *note below
Display, see *note below
CO Reading
O
2
Reading
Dew Point Reading
Channel #1
Channel #3
Channel #4
Channel #2
Visual Alarm 2
A L A R M 2
Air Sample Exhaust
Dew Point Sample Exhaust