5.2 Calibration of the EX-5120 

Calibration is the process of setting the instrument up to read accurately when exposed to a target gas.  The Zero function sets 
the clean air reference point and the Span function sets the sensitivity of the instrument. 

Initial Calibration: 

Wait several minutes for stabilization after supplying power to the 

EX-5120 

sensor/transmitter (S/T) 

before initial calibration.  The S/T has been precalibrated at the factory, and initial field calibration should result in only fine 
tuning to circuit, as well as a way to check that installation is successful.  It is not necessary to open the enclosure to make 
adjustment.  The calibration functions are operated with magnets from outside the enclosure through the 

MENU

 and 

SELECT

 

switches.  Do Not open the S/T unless the area is de-classified. 

Calibration Zero and Span functions are two separate procedures.  They operate independently of each other.  It is 
recommended that the Zero procedure be done prior to the Span procedure.  

ENMET

 Corporation recommends at least 

quarterly calibration of the 

EX-5120 

transmitters. 

Calibration equipment is available from 

ENMET

 Corporation to calibrate the 

EX-5120 

sensor/transmitters.  A calibration 

adapter will have a fitting for the gas cylinder on one side, and a cover to go over the sensor housing on the other.  

Generally, a cylinder of 20.9% Oxygen is used to provide a fresh air reference or Zero point for the calibration.  Another 
cylinder is used to provide the Span reference point for calibration.  Depending on the instrument calibration, the Span gas may 
be the same gas that the instrument is calibrated to display, or it may be another gas, which 

ENMET

 has found to have a similar 

response. 

 

Figure 4: Calibration Adapter EX-5120 Sensor/Transmitter 

5.2.1 Zero Adjust 

A ZERO function should be performed only when the 

EX-5120 

sensor/transmitter is exposed to fresh air.  If the air at the sensor 

is in question, use a cylinder of 20.9% oxygen to provide a clean air reference.  Attach the cylinder to the calibration adapter. 

Enter the maintenance menu by placing the magnet over 

M

ENU

 

switch for 2 to 4 seconds.  See 

Figure 5, 

EX-5120 

Maintenance Menu flow chart. 

The second menu available is the Zero. 

Tap the 

S

ELECT

 switch to perform a Zero. 

 

If the Zero is successful, Cal OK appears on the display and in 1 – 2 seconds, display will change to Span. 
If you wish to Span the sensor Tap the 

S

ELECT

 

switch you are now ready to apply gas.  

Proceed to gas span step 2 

If you wish to Exit the maintenance menu, Tap 

M

ENU

 

switch until Exit is displayed, then tap 

S

ELECT

 

switch to return to the 

instrument Normal Gas Display 

 

If the Zero is Not successful, sensor is outside of safe parameters to be zeroed, the display will read Bad Zero.  Repeat 
Section 5.2.1 Zero Adjust making sure to use a cylinder of 20.9% Oxygen. 

Regulator 

Gas Cylinder 

Select 

Menu 

Output, for Gas 

Input, for Calibration  
or Optional Gas Sampling 

Calibration Cover  

Содержание EX-5120

Страница 1: ...ENMET Corporation PO Box 979 Ann Arbor MI 48106 0979 www enmet com Manual Part Number 80003 092 MCN 13 011 12 20 13 EX 5120 INFRARED 0 100 LEL Sensor Transmitter Manual...

Страница 2: ...tenance Menu 10 5 3 Sensor Replacement 12 6 0 REPLACEMENT PART NUMBERS 13 7 0 WARRANTY 14 List of Figures and Tables Figure 1 EX 5120 Features 2 Figure 2 EX 5120 Mounting Dimensions 3 Figure 3 Termina...

Страница 3: ...ion and after you have received written acknowledgment of the damage from the carrier contact ENMET Customer Service for return authorization and further instructions Have your Purchase Order and Sale...

Страница 4: ...maintenance functions The switch locations are indicated by MENU and SELECT A magnetic field pulse is applied by momentarily putting the end of the magnet in proximity to the switch and then removing...

Страница 5: ...property or personnel when choosing locations of sensor transmitters Also take into consideration environmental factors when deciding on S T location Avoid locations where the S T may be damaged by l...

Страница 6: ...J4 plug then attach to J4 terminal Connect the wires from the sensor to the supplied J8 plug then attach to the J8 terminal See Figure 3 for locations J4 PLUG TERMINAL TO CONTROLLER WIRING Position Fu...

Страница 7: ...n the S T Do Not apply power to the S T without the cover in place Sensor Transmitter Enclosure Cutaway View Printed Circuit Board PCB Display Overlay Display Overlay Screws 2 places Display Overlay S...

Страница 8: ...ansmitter configuration and calibration condition the furthest right character in the display may flash a letter indicating the instrument status Consult ENMET Distributor or ENMET Corp 4 1 Start up 4...

Страница 9: ...splay indicates Alarm 1 Set point Tap the magnet over the MENU button Display indicates Alarm 2 Set point Tap the magnet over the MENU button No Function for the SELECT button in this mode Display ind...

Страница 10: ...de Measurement of target gas Hold the magnet over MENU switch for 2 5 seconds to enter the Maintenance Menu The Power Fault LED will flash Green Red to indicate the EX 5120 is in Maintenance Mode To e...

Страница 11: ...0 9 Oxygen is used to provide a fresh air reference or Zero point for the calibration Another cylinder is used to provide the Span reference point for calibration Depending on the instrument calibrati...

Страница 12: ...il Exit appears then tap the SELECT switch Check span gas and repeat calibration NOTE Some software revisions require the SELECT switch be tapped to accept the signal 7 Calibration is complete 5 2 3 E...

Страница 13: ...set point The MENU switch changes digit indicated by underscore cursor The SELECT switch locks underscored digit and moves to next digit To return to Normal Gas Display Tap MENU switch until EXIT is d...

Страница 14: ...or PC Board Assembly Reassemble the sensor housing After the new sensor has been installed it is suggested to allow the sensor to stabilize for several minutes A Factory calibration must be performed...

Страница 15: ...ropane 0 100 LEL 03056 008 Consult ENMET Distributor or ENMET Corp for additional ranges Regulator Assembly for 17l cylinders 02506 004 Magnet 50030 001 Calibration Sampling Adapter 03700 034 Calibrat...

Страница 16: ...damage caused by the improper use of the product The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise through the use by the purchaser or others...

Страница 17: ...processing delays ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour All instruments sent to ENMET are subject to a minimum...

Страница 18: ...erms K COD Check one K VISA MasterCard______________________ ________ ________ Card number Expiration Card Code K American Express______________________ ________ ________ Card number Expiration Card C...

Отзывы: