background image

EX-5100 C

OMBUSTIBLE

 

 

ENMET 

 

10 

 
 

Calibration: 

Calibration of Standard Gases:

 Calibration of the 

EX-5100

 should be preformed using the target gas unless Methane is 

known to be a possible background hazard.  Reference Table 2 Standard Span Gas 

The 

EX-5100

 was factory calibrated using 50% LEL levels of gases.  Other levels, minimal 20% LEL, are acceptable.  Contact 

ENMET

 prior to use. 

Table 2: EX-5100 Standard Span Gas 

Gas 

Range 

Alarm 1* 

Alarm 2* 

Alarm 3* 

Span Calibration Gas 

Calibration Point 

Methane 

0 – 100 %LEL 

10% LEL 

20% LEL 

50% LEL 

50% LEL Methane 

50% LEL Methane 

Propane 

0 – 100 %LEL 

10% LEL 

20% LEL 

50% LEL 

50% LEL Propane 

50% LEL Propane 

Hydrogen  

0 – 100 %LEL 

10% LEL 

20% LEL 

50% LEL 

50% LEL Hydrogen 

50% LEL Hydrogen 

Butane 

0 – 100 %LEL 

10% LEL 

20% LEL 

50% LEL 

50% LEL Butane 

50% LEL Butane 

*

N

OTE

:

 These internal sensor/transmitter alarms are independent of the 4-20mA Controller alarm point settings. 

C

AUTION

:

 If Methane is a possible hazard, calibration should be done using Methane. 

Calibration of Non-Standard Gases:

 The 

EX-5100

 range is 0 – 100% LEL.  All Non-Standard gas instruments have been 

factory calibrated on LEL of Methane. 

Methane is considered to have a coefficient or relative response of 1.00(one).  Calibration coefficients relative to Methane have 
been generated for over 30 different combustible hydrocarbons and solvents.  A partial list of these gases and vapors are listed 
in Table 3 Non-Standard Gas.  An example of how to use this list of relative response coefficients is shown below. 

Example: Acetone coefficient factor is 2.86 

20%LEL Methane multiplied by the coefficient of 2.86 = 57 

Apply the 20%LEL Methane to the instrument the display should read 57, 57 = 57%LEL Acetone.  See Section 5.2.2 Span Gas. 

C

AUTION

:

 If Methane is a possible hazard, the instrument should be programmed and calibrated for Methane.  Calibration 

should be done using Methane. 

Table 3: Selected EX-5100 Non-Standard Gas 

 

Gas 

 

%LEL 

 

Coefficient 

Calibration Gas 

%LEL Methane 

 

Display Reads 

Acetone 

2.6% 

2.86 

20 

57 

Isobutane 

1.8% 

1.82 

20 

36 

n-Butane 

1.8% 

1.82 

20 

36 

Cyclohexane 

1.3% 

2.22 

20 

44 

Ethane 

3.0% 

1.18 

40 

47 

Ethanol 

3.3% 

2.22 

20 

44 

Ethyl Acetate 

2.2% 

2.86 

20 

57 

Ethylene 

2.7% 

1.54 

40 

62 

Hydrogen 

4.0% 

1.05 

50 

52 

Isopropyl Alcohol (IPA) 

2.2% 

2.86 

20 

57 

Methanol 

6.7% 

1.43 

40 

57 

Methyl Ethyl Ketone(MEK) 

1.9% 

2.86 

20 

57 

n-Pentane 

1.4% 

2.00 

20 

40 

Propane 

2.1% 

1.67 

40 

67 

Toluene 

1.2% 

2.50 

20 

50 

Xylene 

1.1% 

2.50 

20 

50 

In addition to the established coefficients listed in Table 3 above, 

ENMET

 can generate calibration data for most common 

combustible gases and vapors for which a sample can be readily obtained. 

Содержание EX-5100

Страница 1: ...ENMET PO Box 979 Ann Arbor MI 48106 0979 www enmet com Manual part number 80003 096 MCN 416 04 03 09 EX 5100 COMBUSTIBLE Manual...

Страница 2: ...gs 13 5 4 Sensor Replacement 13 6 0 REPLACEMENT PART NUMBERS 14 7 0 WARRANTY 14 List of Figures and Tables FIGURE 1 EX 5100 FEATURES 2 FIGURE 2 EX 5100 MOUNTING DIMENSIONS 3 FIGURE 3 TERMINAL POSITION...

Страница 3: ...spection and after you have received written acknowledgment of the damage from the carrier contact ENMET Customer Service for return authorization and further instructions Have your Purchase Order and...

Страница 4: ...instrument maintenance functions The switch locations are indicated by MENU and SELECT A magnetic field pulse is applied by momentarily putting the end of the magnet in proximity to the switch and the...

Страница 5: ...tters Take into consideration environmental factors when deciding on S T location Avoid locations where the S T may be damaged by liquid immersion excessive heat or other know hazards Also take precau...

Страница 6: ...e supplied J8 plug then attach to the J8 terminal See Figure 3 for locations J4 PLUG TERMINAL TO CONTROLLER WIRING Position Function 1 24 VDC power 2 GND 3 4 20 mA out 4 RS 485 D 5 RS 485 D Contact EN...

Страница 7: ...l Display Mode If intended function Tap the magnet over SELECT switch Tap the magnet over the MENU switch to advance to the Zero procedure For adjusting Zero If intended function Tap the magnet over S...

Страница 8: ...NOTE Software revision may cause variations of display output Example of Display Function The instrument Model EX 5100 The instrument Serial Number The instrument Software Revision IF the right most c...

Страница 9: ...lay indicates Alarm 1 Set point Tap the magnet over the MENU button Display indicates Alarm 2 Set point Tap the magnet over the MENU button No Function for the SELECT button in this mode Display indic...

Страница 10: ...asurement of target gas Hold the magnet over MENU switch for 2 5 seconds to enter the Maintenance Menu The Power Fault LED will flash Green Red to indicate the EX 5100 is in Maintenance Mode To exit t...

Страница 11: ...nd Span functions are two separate procedures They operate independently of each other It is recommended that the Zero procedure be done prior to the Span procedure ENMET recommends at least quarterly...

Страница 12: ...ne Calibration coefficients relative to Methane have been generated for over 30 different combustible hydrocarbons and solvents A partial list of these gases and vapors are listed in Table 3 Non Stand...

Страница 13: ...ociated calibration gas cylinder to the regulator and calibration cover See to Figure 3 4 Open the valve to apply the calibration gas to the sensor 5 Watch for the signal level to stabilize Refer to T...

Страница 14: ...the magnet over the SELECT switch The MENU switch changes digit indicated by underscore cursor The SELECT switch locks underscored digit and moves to next digit See Section 5 2 2 SELECT mA Span MENU S...

Страница 15: ...see Figure 5 for a detailed maintenance menu flow chart Example of Display Function Normal Display Mode Measurement of target gas Hold the magnet over the MENU switch for 2 4 seconds to enter the Mai...

Страница 16: ...means of transportation is requested during the warranty period the customer is responsible for the difference between the most economical means and the expedient mode ENMET shall not be liable for an...

Страница 17: ...ult in processing delays ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour All instruments sent to ENMET are subject to a mi...

Страница 18: ...OD Check one K VISA MasterCard______________________ ________ ________ Card number Expiration Card Code K American Express______________________ ________ ________ Card number Expiration Card Code Name...

Отзывы: