Tophit H CPS471/CPS471D
Hauser
11
Sensor orientation
When installing the sensor, note the flow-past direction of the medium. The ISFET chip should be fixed at an
angle of approx 45° to the flow-past direction (
12). Fixing at the correct angle is very easy because of
the rotable plug-in head.
When installing the sensor in an assembly, use the engraved serial number on the connection head for correct
sensor orientation. The serial number is always located in the same plane as the ISFET chip and the nameplate
(z-y-direction,
a0003893
Fig. 11:
Sensor orientation, front view
1
Serial number
2
Nameplate
a0003891
Fig. 12:
Sensor orientation, 3d view
1
Serial number
3
rotable part of the connection head
4
Medium flow-past direction
5
ISFET chip