Chemistry and process (Ex applications)

Process sterilizibility is no problem due to the wide range of applications for the ISFET pH sensor, not only 
relating to temperature but also to pH. There is only a small range of high pH values connected with high 
temperatures where the sensor is not constantly stable (see "Process"). Media with these characteristics remove 
the isolator oxide from the ISFET chip. As this is the pH and temperature range of CIP cleaning media, the 
ISFET pH sensor should only be used in combination with an automatic retractable assembly.

Benefits of the Topcal fully automatic measurement, cleaning and calibration system:

• CIP cleaning

The sensor built into the retractable assembly is automatically "moved" out of the medium before cleaning. 
In the rinse chamber of the retractable assembly the sensor is cleaned with suitable cleaning solutions.

• Calibration cycles can individually be set.
• Low maintenance costs due to fully automatic cleaning and calibration functions.
• Measuring results are optimally reproducable and the individual value tolerances are very low due to the 

automatic calibration.

a0003880-en

Fig. 6: 

Measuring system with fully automatic measuring, cleaning and calibration system Topcal S

1
2
3
4

Tophit
Cleanfit H CPA475
Mycom S CPM153
Power supply

5
6
7
8

Control unit CPG310
Cleaner, buffer solutions
Steam, water, cleaner
Rinse block

9
10
11

Power cable
Compressed air
Liquids / cleaner

Содержание Tophit CPS471

Страница 1: ...ference system poisoning resistant polyacrylamide free gel Application possible at low temperatures Short response time Constantly high accuracy Sterilisable and autoclavable Longer calibration interv...

Страница 2: ...table sensor circuits use the dependence on the ion selective gate potential to create an output signal proportional to the concentration of the ion type pH selective IsFET The gate isolator serves as...

Страница 3: ...lues 0 09 and 13 86 a0003867 en Fig 3 Comparison of acid and alkaline errors Measurement stability and sensor response time The ISFET response times are very short over the whole temperature range Wit...

Страница 4: ...H value can be worked out from this equation isothermic curve potential change per pH value at a defined temperature The isothermic curves of the ISFET sensor are very close to the theoretical values...

Страница 5: ...mitter and used to calculate the current measured value Storing the calibration data in the sensor allows for calibration and adjustment away from the measuring point The result Sensors can be calibra...

Страница 6: ...sembly e g Cleanfit P CPA471 CPA450 with digital sensor only There are additional accessories available depending on the application Topclean S CPC30 or Topcal S CPC310 automatic cleaning system Exten...

Страница 7: ...tomatic measurement cleaning and calibration system CIP cleaning The sensor built into the retractable assembly is automatically moved out of the medium before cleaning In the rinse chamber of the ret...

Страница 8: ...mmersion assembly Dipfit W CPA111 3 Special measuring cable CPK12 or CYK10 4 Transmitter Liquiline Pharmaceuticals and biotechnology a0003877 Fig 8 Measuring system for pharmaceutical and biotechnolog...

Страница 9: ...ote The cable cors Yellow and White are connected on the sensor side Make sure you comply with the instructions for connecting the sensor wiring diagram in the Operating Instructions of the transmitte...

Страница 10: ...s Longer dry storage of the sensor up to 30 minutes Installation Installation angle ISFET sensors can be installed in any position as there is no liquid internal lead However in case of an overhead in...

Страница 11: ...he rotable plug in head When installing the sensor in an assembly use the engraved serial number on the connection head for correct sensor orientation The serial number is always located in the same p...

Страница 12: ...alibration and operation Normal environment light does not influence the measurement Process Medium temperature depending on pH At high temperatures over a long period of time alkalis irreversibly des...

Страница 13: ...en Fig 14 Pressure and temperature Caution Danger of damage to the sensor Never use the Tophit for applications outside the given specifications Recommended cleaning Depending on the degree of polluti...

Страница 14: ...acc to DIN IEC 751 Plug in head CPS471 ESB TOP68 rotatable CPS471D Memosens rotatable Diaphragm Ceramics sterilizable a0003903 Fig 15 Tophit CPS471 depending on the sensor version a0003904 Fig 16 Sens...

Страница 15: ...l EPDM Application range pH 0 to 14 15 to 135 C 5 to 275 F For Ex and Non Ex applications Product structure CPS471 Product structure CPS471D Shaft length 2 120 mm 4 72 in 4 225 mm 8 86 in 5 360 mm 14...

Страница 16: ...obtained are representative of the ISFET Sensor TopHit type CPS 471 and type CPS 441 Materials in contact with medium are FDA certified Certified acc to the 3 A standard no 74 02 Accessories Transmit...

Страница 17: ...iable process termination parts in contact with medium of PP PEEK or PVDF Technical Information TI345C 07 en Cleanfit H CPA475 Retractable assembly for tank and pipe installation under sterile conditi...

Страница 18: ...sheath max 130 C 266 F HC Cable length 15 m 49 23 ft TPE sheath max 130 C 266 F HD Cable length 20 m 65 64 ft TPE sheath max 130 C 266 F HF Cable length 5 to 20 m 16 41 to 65 64 ft TPE sheath max 130...

Страница 19: ...of e g Memosens CUS31 CUS41 2 wires twisted pair with shield and PVC sheath 2 x 2 x 0 5 mm2 shield Sold by the meter order no 51502543 Junction boxes Junction box VBA With 10 high impedance terminals...

Страница 20: ...TI283C 07 en 04 06 51506685 Printed in Germany FM SGML 6 0 DT...

Отзывы: