8 Troubleshooting
Topclean S
132
Hauser
8.2
Response of outputs to errors
8.2.1
Current output behaviour
If an error occurs in the system, an error current is output at the current output. You can adjust the
value of this error current in the Alarm menu (see page 65).
If you have configured the controller for functioning with current output 2, this output does not
output an error current if an error occurs.
8.2.2
Response of contacts to errors
You can set each individual error message to trigger or not to trigger an alarm (see table on page 125,
editing errors on page 65). In "NAMUR" mode, failure messages always trigger an alarm.
Behaviour with standard setting
Instrument gets hot
Incorrect mains voltage or too high
Compare mains voltage and nameplate
rating.
Power unit defective
Replace power unit.
All 6 red LEDs on the M3G module must
be lit
Incorrect measured pH /
mV value and / or
measured temperature
value
Transmitter module defective (module:
MKP2), please carry out tests and
measures first as described in chap. 8.1.2
Test measuring inputs:
– Connect pH, Ref and PA directly on the
instrument with wire jumpers:
Display must be pH 7.
– Resistance 100
Ω
of Terminals 11 to 12+
13.
Display must be 0 °C.
If test negative: Replace module MKP2,
bushing using the device view on
page 135
Current output,
Current value incorrect
Calibration incorrect:
– Load too high
– Shunt / short-circuit to frame in
current circuit
Test with integrated current simulation,
connect mA meter directly to current
output.
If simulation value incorrect: new module
M3Cx-x
required.
If simulation value correct:
check current circuit for load and shunts
Incorrect operating mode
Check whether 0–20 mA or 4–20 mA is
selected.
No current output signal
Jumper incorrectly set
Change coding.
See chap. 8.4.2
Current output stage defective
(Module: M3CH-x)
Test with integrated current simulation,
connect mA meter directly to current
output.
If test negative:
Replace module M3CH-x
(Check variants, see spare parts list in
chap. 8.3)
Instrument with PROFIBUS
®
interface
PROFIBUS
®
instruments do not have a cur-
rent output.
For information, see "DIAG" / internal
data
Error
Possible cause
Tests and / or remedial action
Equipment, spare parts,
personnel
Instrument status
Alarm relay
Limit value / Controller
Normal operation
picked-up (Fail-safe behaviour)
Corresponding to
configuration and operating
status
Alarm
Dropped out
Voltage-free
Dropped out
Dropped out
Hold
Picked-up
Corresponding to
configuration and operating
status.
Содержание Topclean S CPC30
Страница 154: ...10 Technical data Topclean S 154 Endress Hauser...
Страница 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Страница 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Страница 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Страница 159: ...Topclean S 11 Appendix Endress Hauser 159...
Страница 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Страница 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Страница 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Страница 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Страница 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Страница 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Страница 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Страница 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Страница 176: ...Index Topclean S 176 Endress Hauser...
Страница 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...