Topclean S
8 Troubleshooting
Hauser
129
8.1.2
Process-specific errors
Error
Possible cause
Remedial action
Equipment needed, spare
parts
Instrument
unconfigurable,
Display for code prompt is
9999
Instrument hardware is locked via key-
pad (Keys "CAL" + "DIAG" simultane-
ously = security locking)
Press "MEAS" and "PARAM" simultaneously
to unlock.
Measuring chain zero
point not settable
Reference system poisoned
Test with new electrode.
pH/mV electrode
Diaphragm blocked
Clean diaphragm or grind ceramic
diaphragm.
HCl 3%, file (ceramic diaphragm only,
only file in one direction)
Measuring line broken
pH input on instrument
short-circuit
⇒
Display pH7
Reterminate or use new cable
Unsymmetrical sensor voltage too high
Clean diaphragm or test with different
electrode.
HCl 3%, file (ceramic diaphragm only,
only file in one direction)
Potential matching (PA/PM) Mycom
⇔
Incorrect medium
asymm.: no PM or PM on PE
symm.: PM connection needed
Connection s. chap. 4.1.3
No calibration possible, as
sensor settling time is too
long
ISFET sensors only: liquid film between
semiconductor and reference lead is
interrupted due to drying or cleaning
with compressed air.
Make sure that the liquid film is established
or that the buffer remains in the rinse
chamber for more than 6 min.
Do not use compressed air to clean the
ISFET sensor.
No calibration possible as
measuring chain slope not
adjustable/slope too small
Connection not at high impedance
(humidity, dirt)
Test cable, connector and junction boxes.
pH simulator
Instrument input defective
Directly test instrument.
pH simulator
Electrode aged
Replace electrode.
pH electrode
No calibration possible as
measuring chain slope not
adjustable/no slope
Hairline crack in glass membrane
Replace electrode.
pH electrode
Connection not at high impedance
(humidity, dirt)
Test cable, connector and junction boxes.
pH simulator, see also Chap. 7.3.2
Semi-conductor layer in measuring cable
not removed
Check inner coaxial cable, remove black
layer.
See page 23.
No or slow display change
Electrode soiled
Clean electrode.
Electrode aged
Replace electrode.
New electrode
Electrode defective (reference lead)
Replace electrode.
New electrode
Inner buffer missing
Electrodes with liquid KCl only (CPS41,
CPS441): Check KCl supply (0.8 bar above
medium pressure).
KCl (CPY4-x)
Problem with diaphragm or missing
electrolyte
Permanent, incorrect
measured value
Electrode not immersing or protective
cap not removed
Check installation position, remove
protective cap.
Air cushion in assembly
Check assembly and installation position.
Earth fault at or in the instrument
Test measurement in insulated vessel,
possibly with buffer solution.
Plastic vessel, buffer solutions. Behaviour
when instrument is connected to process?
Shunt due to moisture in cable or plug
Check cable with plug for high impedance.
Replace if necessary.
Hairline crack in glass membrane
Replace electrode.
pH electrode
Instrument in impermissible operating
status (no response on pressing key)
Switch instrument off and on.
EMC problem: If repeated, check
grounding and wire routing
Incorrect temperature
reading
Incorrect sensor connection
Check connections using wiring diagram.
Wiring diagram see page 38
Measuring cable defective
Check cable.
Ohmmeter
Incorrect sensor type selected
Set sensor type on instrument.
Check temperature sensor with
Ohmmeter.
Sensor defective
Check sensor.
Содержание Topclean S CPC30
Страница 154: ...10 Technical data Topclean S 154 Endress Hauser...
Страница 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Страница 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Страница 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Страница 159: ...Topclean S 11 Appendix Endress Hauser 159...
Страница 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Страница 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Страница 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Страница 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Страница 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Страница 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Страница 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Страница 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Страница 176: ...Index Topclean S 176 Endress Hauser...
Страница 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...