14.4 Power supply
Supply voltage
100/240 V AC, 47 - 63 Hz
Power consumption
Max. 60 W
Power supply cable
2 m, Type E+F mains plug pre-installed
14.5 Performance characteristics
Measuring range
TOC (total organic carbon)
Maximum measured error
+/- 0.5 µg/l (ppb) or 1 %, the larger value applies in each case
Limit of detection (LOD)
0.1 µg/l (ppb)
Response time t90
50 s
Number of measuring
channels
1 to 3, depending on the order version
Sample requirement
14 ml/min.
UV reactor
UV reactor with continuous function monitoring
Calibration interval
The device is calibrated on delivery. It is recommended to perform a new calibration after
replacing components in contact with the process, such as the pump hose or UV reactor.
Maintenance interval
• Replacement of calibration solution - before every calibration
• Replacement of pump hose - every 6 months
• Replacement of UV reactor - every 6 months
• Replacement of UV reactor ballast - every 24 - 36 months
• Replacement of pump head - every 36 - 48 months
Maintenance effort
1 hour per month