Commissioning
TOC Analyzer CA78
20
Hauser
A0046940
Warnings
If the TOC and/or conductivity is above the configured limit, the value is displayed in red
digits. In addition, a warning is output via a digital output (optional). The limit can be
shown as a red line in the chart.
A0046943
7
Measurement over range limit
Leak in the system
If there is a leak in the system, the analyzer stops the measurement automatically and
closes valve 1. Once the leak has been found and eliminated, the analyzer starts measuring
again (only if the
Continue after error
option is enabled in the
Settings
menu,
Options 2
tab → 27. The leak sensor in the system must be dried thoroughly beforehand. The
leak sensor is located on the base of the device on the right-hand side.
Содержание TOC Analyzer CA78
Страница 2: ......
Страница 15: ...TOC Analyzer CA78 Mounting Endress Hauser 15 2 Inspect all the hoses for any damage Replace damaged hoses...