TOC Analyzer CA78
Mounting
Hauser
13
264 (10.39)
450 (17.72)
18 (0.70)
18 (0.70)
25 (0.98)
25 (0.98)
1
A0047197
4
Rear of the housing
1
Blind rivet nut
5.2
Mounting the analyzer
L
WARNING
The device is live!
Risk of electric shock!
‣
Do not connect the analyzer to the electricity supply before the installation work is
completed and the media are connected.
‣
Follow the instructions in the "Electrical connection" section.
5.2.1
Mounting sequence
Bench-top mounting
1. Place the analyzer on an even, vibration-free surface.
2. Open the front door of the housing and check the interior structure for visible signs
of damage.
3. Check all the built-in fluid connections. The hoses must not be bent or damaged.
4. Check all the fluid couplings to ensure they are fitted securely (tighten by hand).