Teqwave T
Operation
Hauser
37
9.5
Accessing the measured data
Once the connection has been established to a computer via the USB mass storage medium
or via the operating tool, measured values saved in the measuring device can be deleted from
the measuring device or read out and saved as a .csv file. With the operating tool it is also
possible to create a measured value report for the selected measurement point in PDF
format.
The measuring device can save a maximum of 3000 measured values per measurement
point with a maximum of 50 measurement points per measuring device.
9.5.1
Reading out and/or deleting measured values via the USB
mass storage medium
Spread of malicious software
The computer recognizes the transmitter as a mass storage medium. Because of this, mal-
ware that may be on the computer can potentially spread.
‣
Ensure suitable security measures are taken, such as the use of anti-virus software.
The measuring device cannot perform any measurements while it is connected to the
computer.
1. Connect the measuring device to the computer (switched on) via the USB cable.
2. Switch on the measuring device with the On/Off button.
The computer recognizes the measuring device as a USB mass storage medium.
3. Double-click the USB mass storage medium on the computer to open it.
4. Double-click the "Data" folder to open it.
The folder contains a .csv file with the measured values for each measurement point.
5. If necessary, save the desired .csv file to the computer using the Copy function or
double-click the file to open it.
6. If necessary, delete the desired .csv file using the Delete function.
9.5.2
Reading out and/or deleting measured values via the
operating tool
1. Select the menu "File"
→
"Save data from selected measurement point" or "Save data from
all measurement points".
The "Select storage location" window is displayed.
2. Select the location for storing the .csv file containing the measured values.
The operating tool saves the .csv file in the selected directory.
3. Select the menu "Device con
fi
guration"
→
"Measured values"
→
"Delete (current)" or
"Delete (all)".
The operating tool deletes the .csv files containing the measured values from the
measuring device.
9.5.3
Generating a measured value report via the operating tool
Making initial report settings
1. Select the menu "Report"
→
"Settings".
The "Report configuration" window appears.
2. Make the desired settings.
3. Click "Save" to save the settings made.
The operating tool uses the settings when generating the report.
NOTICE
Содержание Teqwave T
Страница 54: ...www addresses endress com ...