Stamolys CA71SI
6
Hauser
Performance characteristics
Time between two
measurements
t
mes
= reaction time + rinse time + waiting time + rinse again time + filling time + sampling time +
reagent refusal time (min. waiting time = 0 min)
Maximum measured error
1 to 200
μ
g/l:
±2 % of measuring range end
there of 5 to 20
μ
g/l:
±2
μ
g/l
50 to 5000
μ
g/l:
±2 % of measuring range end
Measuring interval
t
meas
to 120 min
Reaction time
• SI-A:
90 s
• SI-B:
120 s
Sample requirement
20 ml (0.68 fl.oz.) per measurement
Reagent requirement
SI-A
3 x 0.288 ml (0.01 fl.oz.)
0.86 l (0.23 US.gal) per reagent per month with 15 minute measuring interval
SI-B
3 x 0.18 ml (0.006 fl.oz.)
0.52 l (0.14 US.gal) per reagent per month with 15 minute measuring interval
Calibration interval
0 to 720 h
Rinse interval (SI-B only)
0 to 720 h
Rinse again time
30 s
Filling time
SI-A
24 s
SI-B
17 s
Sample refusal
t
refusal
= 30 s (SI-A)
t
refusal
= 0 s (SI-B)
Maintenance interval
6 months (typical)
Servicing requirement
15 minutes per week (typical)
Environment
Ambient temperature
5 to 35
˚
C (40 to 95
˚
F), avoid strong fluctuations
Humidity
below the condensation limit, installation in usual, clean rooms
outdoor installation only possible with protective devices (customer supplied)
Ingress protection
IP 54