Technical Information

Stamolys CA71SI

Silicate Analyzer
Compact photometric analysis system for the silicate measurement in 
ultrapure water and boiler feedwater

Application

• Ultrapure water
• Boiler feedwater
• Steam and condensate analysis
• Reversed osmosis
• Demineralizers

Industries:
• Pharmaceutical industry
• Power stations

Your benefits

Stamolys CA71SI measures the silicate content quasi-
continuously to assure a constant optimum water quality. This is 
of special advantage in the performance monitoring of ion 
exchange and reversed osmosis systems. The analyzer replaces 
frequent manual sampling and exactly determines the silicate 
break through of the system.

• Stainless steel or glass-fibre reinforced carbon housing 

available

• Two channel version available
• Measured value storage using integrated data logger
• Automatic calibration and self-cleaning
• Free selectable measuring, cleaning and calibration intervals

Содержание Stamolys CA71SI Series

Страница 1: ...olys CA71SI measures the silicate content quasi continuously to assure a constant optimum water quality This is of special advantage in the performance monitoring of ion exchange and reversed osmosis systems The analyzer replaces frequent manual sampling and exactly determines the silicate break through of the system Stainless steel or glass fibre reinforced carbon housing available Two channel ve...

Страница 2: ...ight is measured by the detector and converted to an electrical signal In the analyzer the respective concentration is calculated from the proportion of light absorbed by the sample Silicate With 18 silicon is the second most frequent element of the earth crust It occurs chemically bonded as silicate or silicon dioxide in a variety of minerals From these minerals silicate or silicic acid is washed...

Страница 3: ...ent This results in a constantly high flow rate The high flow rate generates the self cleaning effect Therefore mechanical drives for the generation of a flow at the filter surface are not necessary Analyzer without Endress Hauser sample conditioning Before analysis the sample has to be conditioned homogeneous and to be transported to an external or to the delivered collecting vessel Measuring sys...

Страница 4: ...ignal on alarm Contacts 2 limit contacts per channel 1 system alarm contact optional end of measurement with two channel version display of channel no available Load max 500 Ω Load capacity 230 V 115 V AC max 2 A 30 V DC max 1 A Data interface RS 232 C Data logger 1024 data pairs per channel with date time and measured value 100 data pairs with date time and measured value for calibration factor d...

Страница 5: ...D GN YE GN YE 80 80 75 75 70 70 63 63 57 57 39 39 27 27 21 21 15 15 64 64 58 58 40 40 28 28 22 22 16 16 65 65 79 79 74 74 69 69 62 62 56 56 38 38 26 26 20 20 14 14 78 78 73 73 68 68 61 61 55 55 37 37 25 25 19 19 13 13 77 77 72 72 67 67 60 60 54 54 36 36 24 24 18 18 12 12 76 76 71 71 66 66 59 59 53 53 35 35 23 23 17 17 11 11 RELAIS Relay max Last load 2A bei at 115 230V AC 1A bei at 30V DC RELAIS R...

Страница 6: ...88 ml 0 01 fl oz 0 86 l 0 23 US gal per reagent per month with 15 minute measuring interval SI B 3 x 0 18 ml 0 006 fl oz 0 52 l 0 14 US gal per reagent per month with 15 minute measuring interval Calibration interval 0 to 720 h Rinse interval SI B only 0 to 720 h Rinse again time 30 s Filling time SI A 24 s SI B 17 s Sample refusal trefusal 30 s SI A trefusal 0 s SI B Maintenance interval 6 months...

Страница 7: ...ample flow rate min 5 ml 0 17 fl oz per min Consistence of the sample low solid content 50 ppm Sample inlet Unpressurized pH value of the sample pH 3 not buffered Mechanical construction Design dimensions Analyzer stainless steel version SI A and SI B a0001361 Stainless steel version ...

Страница 8: ...9 57 200 7 87 120 4 72 326 12 8 330 13 0 110 4 33 Ø 27 1 06 Ø 16 0 63 Ø 27 1 06 120 4 72 555 21 8 465 18 3 500 19 7 7 11 0 28 0 43 M 6 Ø 26 1 02 mm inch 38 1 5 60 2 36 100 3 94 140 5 51 160 6 30 250 9 84 125 4 92 165 6 50 Ø 21 0 83 Ø 26 1 02 Ø 21 0 83 119 4 69 78 3 07 47 1 85 648 25 5 360 14 2 436 17 2 243 9 57 200 7 87 120 4 72 326 12 8 Ø 16 0 63 Ø 27 1 06 120 4 72 555 21 8 465 18 3 500 19 7 7 11...

Страница 9: ...imensions 90 x 90 x 215 mm 3 54 x 5 54 x 8 46 inch The number of bottles varies from 2 to 5 depending on the analyzer version For these versions the outlet pipe must be installed right of the analyzer The outlet pipe must be mounted to a wall so that the sample outlet hoses from the photometer have a gradient of 5 to 10 If neccessary extend the hoses RS 232 RS 232 381 15 0 351 13 8 325 12 8 8 0 31...

Страница 10: ...zer outlet 130 5 12 17 0 67 10 0 39 145 5 71 79 3 11 120 4 72 140 5 51 115 4 53 Ø 34 25 1 34 0 98 3 x Ø 28 20 1 1 0 79 4 x Ø 6 6 0 26 27 1 06 60 2 36 4 x Ø 5 5 10 4 0 22 0 41 Ø 56 2 2 Ø 50H7 1 97H0 28 M20x1 5 Ø 24 0 94 70 2 76 100 3 9 80 3 15 mm inch 48 1 89 42 1 65 70 2 76 60 2 36 GFR housing Stainless steel housing Without housing approx 28 kg 62 lbs approx 33 kg 73 lbs approx 25 kg 55 lbs Housi...

Страница 11: ... the product by affixing the 4 symbol Test reports Quality certificate Depending on the order code you receive a quality certificate With the certificate the manufacturer confirms compliance with all technical regulations and the successful individual testing of your product Connection Hose ID 16 mm 0 63 inch Max length of closed loop 1 m 3 3 ft Open outlet downgrade installed No combination of se...

Страница 12: ...ate a complete and valid order code on the Internet using the configurator tool Measuring range A Measuring range 1 0 200 μg l SiO2 B Measuring range 50 5000 μg l SiO2 Y Special version acc to customer s specification Sample transfer 1 Sample transfer from one measuring point one channel version 2 Sample transfer from two measuring points two channel version Power supply 0 Power supply 230 V AC 50...

Страница 13: ...m 0 13 inch x 0 13 inch x 0 13 inch An interference suppressor for the current output A nozzle for outlet pipe ID 16 mm 0 63 inch A hose clamp Two pipe clamps version without housing only A screwed socket for the outlet pipe Four edge covers A roll of PTFE strip A quality certificate Operating Instructions English SI B The scope of delivery comprises An analyzer with mains plug A cleaning injector...

Страница 14: ...747 V02AAE Collecting vessel for sampling from pressurized systems results in an unpressurised continuous sample stream Collecting vessel without level measurement order no 51512088 Collecting vessel with level monitoring for pure and ultrapure water level monitoring with polypropylene float lever Order no C A061019 50 Maintenance kit CAV740 maintenance kit for CA71 Pump hoses Valve hoses Hose con...

Страница 15: ...Stamolys CA71SI Endress Hauser 15 ...

Страница 16: ...TI00364C 07 EN 14 12 71161856 FM9 ...

Отзывы: