Stamolys CA71CODcr
Technical data
Hauser
51
10.5
Installation
Mounting location
Installation on a vibration-free wall
10.6
Environment
Ambient temperature
10 to 35
˚
C (50 to 95
˚
F), avoid strong fluctuations
Humidity
below the condensation limit, installation in usual, clean rooms
outdoor installation only possible with protective devices (customer supplied)
Ingress protection
IP 43
10.7
Process
Sample temperature
5 to 40
˚
C (40 to 100
˚
F)
Sample flow rate
min. 5 ml (0.17 fl.oz.) per min
Consistency of the sample
aqueous and homogenized
Sample inlet
Unpressurized
10.8
Mechanical construction
Design, dimensions
see chapter "Installation"
Weight
approx. 32 kg (71 lbs)
Material
Housing:
Front windows:
Endless hose:
Pump hose:
Valves:
Glass-fibre reinforced carbon (GFR)
Polycarbonate
C-FLEX, NORPRENE, PTFE, PFA
TYGON