Installation
StamoLys CA 71 SI
12
Hauser
8.
Connect the tubes from canisters containing reagents, standard and cleaning
agents to the following nozzles:
!
Note!
The pressure of the hose cassette is set at the factory such that the sample and reagent
are fed in without bubbles.
Only change the pressure if the factory setting does not meet your requirements. The
setting is changed by turning the adjustment screw using a 2.5 mm Allen key.
3.4
Installation examples
3.4.1
CAT 430 or customer-specific ultra filtration and two
CA 71 analyzers
Canister
Hose designation (mark)
Sample
Reagent 1
Reagent 2
Reagent 3
Standard solution
Cleaner
P
SI1
SI2
SI3
S
R
•
Permeate can contain air bubbles
(CAT 430) or is free of bubbles
(customer-supplied ultrafiltration)
•
Distance between the analyzers as short
as possible: sampling line between
T-piece and the second analyzer
(Fig. 10, item 2) shorter than 1.5 m
•
Cross-section of sampling line
ID 3.2 - 4 mm
•
only one sample receiver required
!
Note!
Ensure that there is always sufficient
sample available for both analyzers.
Observe this when selecting maintenance
intervals for CAT 430 and when setting the
buffer volume on the collecting vessel.
C07-CA71xxx-11-08-00-xx-007.eps
Fig. 10:
Installation example
1
Sample from CAT 430
2
Sampling line
3
Sample receiver overflow
4
T-piece
5
Collecting vessel
1
4
5
CA 71
CA 71
2
3