Installation
StamoLys CA 71 SI
10
Hauser
One-channel version
Two-channel version
•
Due to the ordered version are oneor two collecting vessels (with or without level
measurement) in the scope of delivery.
•
Level measurement is only for one channel possible.
•
Only one collecting vessel can be mounted at housing. The second is to be placed
nearby the analyzer.
Adjusting the level measurement (one-channel version only)
Adjust the conductive level measurement due to the number of connected analyzers.
1.
In dependence of the application, mount the right adjusting pin or no adjusting pin
(Fig. 6 and Fig. 7, position 2).
2.
Pull the marked pipe (position 3) downwards due to your application (1, 2 or 3
analyzers).
E+H Collecting vessel
(at CA 71, with or without level measurement)
Connection
hose ID 3.2 mm (0.13")
Customer collecting vessel
Connection
hose ID 1.6 mm (0.06")
Max. distance from collecting vessel to analyzer
1 m (3.28 ft)
Max. height difference from collecting vessel to
analyzer
0.5 m (1.64 ft)
C07-CA71xxx-11-08-00-xx-004.eps
Fig. 6:
One Analyzer
1
M 3x12 (0.47")
2
M 3x35 (1.38")
3
Mark 1
C07-CA71xxx-11-08-00-xx-003.eps
Fig. 7:
Two analyzers
1
M 3x12 (0.47")
2
M 3x20 (0.79")
3
Mark 2
C07-CA71XXx-11-08-02-xx-001.eps
Fig. 8:
Three analyzers
1
M 3x12 (0.47")
3
Mark 3
1
2
3
1
2
3
1
3