51512332
Valid of:
Software version 5.2

StamoLys CA 71 SI

Silicate Analyzer

Operating Instructions

Содержание Stamolys CA71CL

Страница 1: ...Quality made by Endress Hauser ISO 9001 BA 364C 07 en 07 03 51512332 Valid of Software version 5 2 StamoLys CA 71 SI Silicate Analyzer Operating Instructions...

Страница 2: ......

Страница 3: ...eration 20 5 4 Calibration 28 6 Commissioning 30 6 1 Function check 30 6 2 Switch on 30 7 Maintenance 32 7 1 Maintenance schedule 32 7 2 Replacing reagents 33 7 3 Replacing pump hoses 34 7 4 Replacing...

Страница 4: ...cular the analyzer is intended for steam and condensate analysis reversed osmosis demineralisers Any other use than the one described here compromises the safety of persons and the entire measuring sy...

Страница 5: ...n it cleaned to the Endress Hauser sales centre responsible Please use the original packaging if possible Please enclose the completed Dangerous Goods sheet copy the second last page of these Operatin...

Страница 6: ...Sample transfer from two measuring points two channel version Power supply 0 Power supply 230 V AC 50 Hz 1 Power supply 115 V AC 60 Hz Collecting vessel for up to 3 analyzers A Without collecting ves...

Страница 7: ...output 4 edge covers a quality certificate Operating Instructions english Note Reagents please order separately with analyzer version CA 71 XX XXXXXX1 With versions CA 71 XX XXXXXX2 and CA 71 XX XXXXX...

Страница 8: ...eck that the scope of delivery is complete and agrees with your order and the shipping documents The packaging material used to store or to transport the product must provide shock protection and humi...

Страница 9: ...11 08 00 xx 001 eps Fig 4 Collecting vessel at analyzer optionally 1 Ventilation 2 Sample inlet from sampling 3 Collecting vessel 4 Electrical connections 5 Analyzer sample inlet C07 CA71xxx 06 08 02...

Страница 10: ...cation mount the right adjusting pin or no adjusting pin Fig 6 and Fig 7 position 2 2 Pull the marked pipe position 3 downwards due to your application 1 2 or 3 analyzers E H Collecting vessel at CA 7...

Страница 11: ...valve hoses V1 Valve 1 V2 Valve 2 1 To the pump 2 Y piece connecting hose to valve 1 behind it 3 Hose valve 2 front standard 4 Hose valve 2 back cleaning agent 5 Hose valve 1 front sample 6 Secure th...

Страница 12: ...Sample Reagent 1 Reagent 2 Reagent 3 Standard solution Cleaner P SI1 SI2 SI3 S R Permeate can contain air bubbles CAT 430 or is free of bubbles customer supplied ultrafiltration Distance between the a...

Страница 13: ...s short as possible sampling line between T piece and the second analyzer Fig 11 item 5 shorter than 1 5 m Cross section of sampling line ID 3 2 4 mm one sample receiver each without level measurement...

Страница 14: ...onnel must have read and understood the instructions in this manual and must adhere to them Ensure that there is no voltage at the power cable before beginning the connection work 4 1 1 Quick wiring g...

Страница 15: ...not ready To start measurement the voltage must stay at 0 V for at least 5 seconds Function Designation Terminal one channel Terminal two channels Mains L 3 3 N 2 2 PE 1 1 Alarm value 1 channel 1 COM...

Страница 16: ...l version place the cable cores of all cables data cables to analog output 1 and to analog output 2 through the interference suppressor 4 2 2 One channel version C07 CA71xxx 03 08 00 xx 001 eps Fig 13...

Страница 17: ...ult Retrieves error message using operation menu Channel or measurement end1 1 Alternative selection Displays active channel Displays Measurement finished 5 s Analog output l 1 channel 1 0 or 4 mA mea...

Страница 18: ...g commands can be sent from the PC in order to read out data D Data memory Measured values C Data memory Calibration factors S Setup parameter entry configuration F Frequency current Connection Termin...

Страница 19: ...e check the terminal assignment for analogue output 1 or 2 Device status and specifications Note Is the analyzer or cable externally undamaged Visual inspection Electrical connection Note Does the sup...

Страница 20: ...rial interface RS 232 4 Operating keys and control LEDs 5 3 Local operation The operating keys and the integrated indicator LEDs have the following functions RS 232 ENDRESS HAUSER STAMOLYS CA 71 M CE...

Страница 21: ...as not yet expired Note Measuring flashes when the analyzer is ready for the next measurement but has not yet received the enable signal from the sample collector or the sample conditioning unit Selec...

Страница 22: ...oncentration of the standard see PARAMETER ENTRY calibration solution The deviation results from factors such as reagent ageing ageing of constructive components etc The calibration factor compensates...

Страница 23: ...only Setting for whether contact for alarm value 1 channel 2 works as NO current or NC current contact Alarm value AV 2 2 Two channel version only Setting for whether contact for alarm value 2 channel...

Страница 24: ...teresis 2 of alarm value Alarm value AV 2 1 SI A 1 200 g l 200 g l SI B 100 5000 g l 5000 g l Concentration threshold value limit relay 2 channel 1 differential hysteresis 2 of alarm value Alarm value...

Страница 25: ...n interval of 48 72 h Calibration solution SI A 1 200 g l 500 g l SI B 100 5000 g l 500 g l Concentration of the calibration standard Select a standard whose concentration is in the upper third of the...

Страница 26: ...y to enter the menu Option Display Info Pumps and valves Virtual switching board Various valve and pump combinations can be selected The setting options are Valve 1 P sample or S standard Pump 1 and p...

Страница 27: ...n ASCII format via the serial interface For this the receiving end PC must be configured like this 9600 N 8 1 To send data the receiving end PC must send the ASCII character 81 Shift D Clear data This...

Страница 28: ...sub menu 5 4 2 Interferences Measuring range Concentration mg l Frequency Hz Silicate lower measuring range SI A 1 200 g l 0 000 0 020 0 040 0 060 0 080 0 100 0 120 0 140 0 160 0 200 0 251 515 831 10...

Страница 29: ...e which lies in the past 6 Press F to accept the value and then press Htwice to return to the main menu 7 Press Fagain This takes you back to measuring mode The calibration is now carried out automati...

Страница 30: ...nfigured and calibrated the measuring cycle starts automatically Entering parameters is no longer necessary To perform initial start up or to readjust the device parameters proceed as follows 1 Plug i...

Страница 31: ...R or S with V and then switch the P1 sample pump on Leave the pump running until you can tell that there is cleaning agent at the T connector Switch P1 back off 7 Now switch the valves to passage for...

Страница 32: ...ekly Check and note calibration factor for service purposes Move valve hoses into their position and spray with silicone extends the service life CONFIGURATION every 2 weeks Check concentration of the...

Страница 33: ...ee Trouble shooting instructions the reagents are too old the reagents have been spoilt by incorrect storage conditions or environmental influences Reagents check a Check the standard solution concent...

Страница 34: ...m the connecting nipples on the pump s 3 Remove the reagent hoses from the reagent canisters in order to prevent contamination of the reagents 4 Loosen the hose cassette Now you can remove the hoses 5...

Страница 35: ...left of the mixer 4 Remove the old mixer from the clip and insert the new one 5 Connect the new mixer to the hoses again 6 Attach the photometer housing and screw it down 7 After installation refill...

Страница 36: ...el SS 1 4301 AISI 304 with a lint free cloth and Glittol RG 10 51 GFR housing with a damp cloth or with tenside based alkaline cleaning agent 7 8 Placing out of service You must place the analyzer out...

Страница 37: ...O2 Si order no CAY642 V10C10AAE Note Please note the disposal instructions in the safety data sheets of the reagents 8 3 Hose cleaner Cleaning agent alkaline 100 ml order no CAY746 V01AAE Cleaning age...

Страница 38: ...ngly PARAMETER ENTRY Calibration solution or replace the standard Reagents contaminated or aged Simple check Add about 20 ml of standard solution with about 1 ml of reagent in a beaker If it does not...

Страница 39: ...methods Detect interfering substances see Technical Information Measuring principle possibly use sample conditioning Standard is dosed to the sample Check the valves and valve settings Replace valve...

Страница 40: ...date Interval not expired or 0 h Change parameters Leak on Photometer Device or discharge blocked Remove blockage E H Service Blockage deposits in device Water hardness Lime deposits can by flushing w...

Страница 41: ...w with sample collector 51504114 122 Hose made of C Flex 6 4 mm 51504115 123 Hose made of C Flex 1 5 mm 51512535 130 Hose adapter 1 6 mm x 1 6 mm 51506495 131 Hose adapter T 1 6 mm x 1 6 mm x 1 6 mm 5...

Страница 42: ...arm Contacts 2 limit contacts per channel 1 system alarm contact optionally end of measurement with two channel version display of channel no available Load max 500 Serial interface RS 232 C Load capa...

Страница 43: ...e time selectable from 20 to 300 s standard 80 s Rinse again time 30 s Filling time 15 s Maintenance interval 6 month typical Servicing requirement 15 minutes per week typical Ambient temperature 5 40...

Страница 44: ...sample Deutsch English Fran ais Italiano Nederlands Suomi Magyar Polski Error messages Pumps and valves Frequency Mixture Measured values Output Clear data Measuring range start Alarm value 1 Calibra...

Страница 45: ...ys CA 71 SI Appendix Endress Hauser 45 Analog output Alarm value 1 Alarm value 2 Error contact act Date time Frequency offset Standard solution Flushing interval Flushing time 1st Flushing Calibrate o...

Страница 46: ...s Suomi Magyar Polski Error messages Pumps and valves Frequency Mixture Measured values Output Clear data Measuring range start1 Calibration Measuring Flushing Waiting PARAMETER ENTRY AUTO MEASURING L...

Страница 47: ...m value 1 2 Calibration solution 1st Flushing act Date time Frequency offset Calibrate offset Error contact Alarm value 2 2 Alarm value 2 1 Analog output 2 Flushing interval Flushing time n Channel 1...

Страница 48: ...ctive 1 l SI1 SI2 SI3 each CAY641 V10AAE Cleaning agent 1 l CAY642 V10C05AAE Standard solution 0 5 mg l SiO2 Si CAY642 V10C10AAE Standard solution 1 0 mg l SiO2 Si Chemicals for Ultra filtration Quant...

Страница 49: ...se ID 6 4 mm 51504115 123 1 m C Flex hose ID 1 5 mm 51512535 130 10 Hose adapter 1 6 mm x 1 6 mm 51506495 131 10 Hose adapter T 1 6 mm x 1 6 mm x 1 6 mm 51506490 132 10 Hose adapter 3 2 mm x 3 2 mm 51...

Страница 50: ...output 0 20 mA 4 20 mA AV 1 normally closed normally open AV 2 normally closed normally open Fault signal normally closed normally open Frequency offset Hz Ground line demineralised water without reag...

Страница 51: ...gain Filling time Reaction time U min K floating mean N Points C1 mg l g l F 1 Hz C2 mg l g l F 2 Hz C3 mg l g l F 3 Hz C4 mg l g l F 4 Hz C5 mg l g l F 5 Hz C6 mg l g l F 6 Hz C7 mg l g l F 7 Hz C8 m...

Страница 52: ...ith pressurised water disposable syringe Remove hose holder unit from the pump Done KW 1 KW 3 KW 5 KW 7 KW 9 KW 11 KW 13 KW 15 KW 17 KW 19 KW 21 KW 23 Date Done KW 25 KW 27 KW 29 KW 31 KW 33 KW 35 KW...

Страница 53: ...y symbols 5 Incoming acceptance 8 Input 42 Installation 4 8 11 Examples 12 Interference suppressor 16 Interferences 28 L LANGUAGE 26 M Main menu 21 Maintenance 32 Schedule 32 Mechanical construction 4...

Страница 54: ...eening 16 Serial interface 18 SERVICE 26 Signal connection 16 Spare parts 40 Standard solution 37 Storage 8 Switching contacts 17 Switch on 30 Symbols 5 T Technical data 42 43 Terminal assignment 15 T...

Страница 55: ...health o biologically hazardous o inflammable o safe Please mark the appropriate warning hints 5HDVRQ IRU UHWXUQ BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB BBBBBBBBBB...

Страница 56: ...ta D C Colsein Ltda Tel 01 2 36 76 59 Fax 01 6 10 78 68 Costa Rica San Jose Euro Tec S A Tel 2 20 28 08 Fax 2 96 15 42 Ecuador Quito Insetec Cia Ltda Tel 02 2 26 91 48 Fax 02 2 46 18 33 El Salvador Sa...