SS2100i-1 and SS2100i-2 Analyzer
4
–
8
4900002288 rev. B 1-13-21
3.
Calculate the new K
p
(K
p
= C
p
/R
p
).
4.
Calculate the Standard Deviation of the mean C
p
5.
Calculate the Validation Allowance as 3
/C
p
* 100.
6.
Press
#2
, enter the Service password and press the
*
key.
7.
Continue pressing the
*
key until the Validation Allowance
parameter displays.
8.
Enter the Validation Allowance calculated above and press the
*
key.
9.
Continue pressing the
*
key until the Val Perm Constant K
p
parameter displays, and enter the new K
p
calculated above.
10.
Enter the new K
p
calculated above and press the
*
key.
11.
Press
#1
to return to Normal mode.
12.
Press
#7
to ensure that the analyzer will continue to provide the
correct Validation Concent/- the Validation Allowance.
Connecting the Validation Gas (Optional)
For systems with an optional manual validation port, an appropriate validation
gas source will need to be connected to the SCS.
To connect the validation source
1.
Determine appropriate tubing route from the customer-supplied
validation gas source to the SCS.
2.
Run stainless steel tubing from the validation source (regulated to
the specified pressure) to the “Validation Gas” supply port.
a.
Bend tubing using industrial grade benders, check tubing fit to
ensure proper seating between the tubing and fittings.
b.
Fully ream all tubing ends.
c.
Blow out the lines for 10 to 15 seconds with clean, dry nitrogen
or air prior to making the connection.
3.
Connect the validation source tube to the SCS using the 1/4”
stainless steel compression-type fittings provided.
4.
Tighten all new fittings 1-1/4 turns with a wrench from finger tight.
For connections with previously swaged ferrules, thread the nut to
the previously pulled up position, then tighten slightly with a wrench.
Secure tubing to appropriate structural supports as required.
5.
Check all connections for gas leaks.
Do not exceed 10 PSIG (0.7 barg) in sample cell. Damage to cell
may result.
Содержание SpectraSensors SS2100i-1
Страница 2: ......
Страница 6: ...SS2100i 1 and SS2100i 2 Analyzer iv 4900002288 rev B 1 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 28: ...SS2100i 1 and SS2100i 2 Analyzer 3 16 4900002288 rev B 1 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 38: ...SS2100i 1 and SS2100i 2 Analyzer A 2 4900002288 rev B 1 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 46: ...SS2100i 1 and SS2100i 2 Analyzer Index 2 4900002288 rev B 1 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 47: ...www spectrasensors com contact 4900002288 Rev B ...