Function codes
Available public functions are:
•
3: Read Holding Registers
•
4: Read Input Registers
•
16: Write Multiple Registers
Please note that function codes 3 and 4 are fully interchangeable as they
behave in the same way.
Data formats
Float
The Float refers to the floating point according to IEEE 754 (Single Precision).
This format requires two registers obtaining 32 bit where each register contains
the high byte in its first bit.
For example, if the float value is 20.56 (int32), represented as 0x41A47AE1
(hexaint32), it is written in two consecutive registers, where the first register
is 3499. Therefore, the value has to be transmitted in the following way:
Int32
All int32 values are 32 bit-wide integer values. The example given in the
previous section applies here as well.
Character
The definition is as follows:
•
8 bit ASCII-Code table according to ISO-8859-1 (Latin-1 Western
European)
Function code 16 may be used with broadcast (device ID = 0).
Codes 3 and 4 may not be used with broadcast.
Register 3499, high byte
0x7A
Register 3499, low byte
0xE1
Register 3500, high byte
0x41
Register 3500, low byte
0xA4
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...