Solicap M FTI55, FTI56
Installation
Hauser
25
Silo roof
Ensure that the silo roof is of a sufficiently stable construction.
High tensile forces may occur when material is being extracted, particularly in the case of heavy and
powdery bulk solids which have a tendency to form buildup.
Coarse-grained bulk solids
In silos with extremely coarse-grained or extremely abrasive bulk solids, the use of a Solicap
M FTI55 or FTI56 is recommended only for maximum detection.
Distance between the rope probes
To rule out mutual probe interference, you must maintain a minimum distance of 0.5 m between
the rope probes. This also applies if you are installing several Solicap M units in adjacent silos with
nonconductive walls.
In the event of condensation:
Use the Solicap M with inactive length.
The inactive length (
A
) prevents moisture and buildup forming between the active part of the probe
and the silo roof.
Or:
To reduce the effects of condensation (
B
) and buildup, the threaded coupling (length: max. 25 mm)
must project into the silo.
Heat insulation reduces condensation and therefore buildup on the steel plate.
A
TI418Fen15
Silo with walls that conduct electricity
B
TI418Fen14
Silo with concrete walls
Active part
of probe
Steel plate
Steel plate;
connected to the
reinforcing steel
Heat insulation
Recommended with condensation: