10.4
Level measurement with automatic calibration correction: two capacitance probes
Start
Chapter 4, Section 4.1
- Constants for EC 37 Z/EC 47 Z or
EB 17 Z/EB 27 Z
- level is displayed at V0H0
Calibration
Chapter 8, Section 8.4
- After calibration level/volume is
displayed at V0H0
Set analogue output signal (optional)
Chapter 6
- Enter settings in the units selected
during calibration or linearization
Set relay parameters (optional)
Chapter 7
- Both relays follow channel 1
Lock parameter matrix (optional)
Chapter 4, Section 4.6
Reset parameters:
V9H5
Enter sensor constants
zero frequency
V3H5
sensitivity
V3H6
Set switching delay
V8H2
Enter sensor position
V8H3
Select sensor type
V8H4
1 = Capacitance probe
Calibrate
V8H5
0 = probe uncovered
1 = probe covered
Set switching point
V8H6
1 = 1Hz, ca. 5 pF (default)
Select mode
V8H0
5 = Calibration correction
Press »E« to activate!!
Assign relay 2
V1H9
1 = channel 1
Set switch point relay 2
V1H5
Set fail-safe mode
V1H6
0 = Minimum
1 = Maximum
Set switching hysteresis V1H7
Set status at fault
V1H8
0 = de-energises
1 = depends on V0H7
Lock parameter matrix
V8H9
Level or volume
V3H0
Correction factor to 1
V8H7
»Empty« calibration
V0H1
Correction factor to 1
V8H7
»Full« calibration
V0H2
Select output range
V0H3
0 = 0
…
20 mA/0
…
10 V,
1 = 4
…
20 mA/2
…
10 V
Set 0/4 mA value
V0H5
Set 20 mA value
V0H6
Set output at fault
V0H7
0 = –10% (–2 mA/–1 V)
1 = +110% (+22 mA/11 V)
2 = last measurement
Set switch point relay 1
V1H0
Set fail-safe mode
V1H1
0 = Minimum
1 = Maximum
Set switching hysteresis V1H2
Set status at fault
V1H3
0 = de-energises
1 = depends on V0H7
Channel 1, level measurement
Silometer FMC 671 Z/676 Z
Chapter 10: Quick programming guide
59