9 Trouble-shooting
Proline Prosonic Flow 90
92
Hauser
9.9
Replacing the device fuse
#
Warning!
Risk of electric shock. Exposed components carry dangerous voltages. Make sure that the power
supply is switched off before you remove the cover of the electronics compartment.
The main fuse is located on the power unit board (Fig. 52).
The procedure for replacing the fuse is as follows:
1.
Switch off power supply.
2.
Remove power unit board
→
3.
Remove protective cap (1) and replace the device fuse (2).
Use only fuses of the following type:
– Power supply 20...55 V AC / 16...62 V DC
→
2.0 A slow-blow / 250 V;
5.2 x 20 mm
– Power supply 85...260 V AC
→
0.8 A slow-blow; 5.2 x 20 mm
– Ex-systems
→
see appropriate Ex documentation
4.
Assembly is the reverse of the disassembly procedure.
"
Caution!
Use only original Hauser replacement parts.
A0001148
Fig. 52:
Replacing the device fuse on the power unit board
1
Protective cap
2
Device fuse
1
2
Содержание Prosonic Flow 90
Страница 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Страница 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Страница 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Страница 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Страница 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Страница 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Страница 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Страница 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Страница 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...