Replacing cover

Promass 80F, 83F, 84F Exd

Hauser

2

1

Safety Instructions

• Comply with national regulations governing mounting, electrical installation, 

commissioning, maintenance and repair procedures. (Available sources, 
e.g. series of standards IEC 60364 and IEC 60079-14/-17/-19).

• Requirements with regard to specialized technical staff for the mounting, 

electrical installation, commissioning, maintenance and repair of the 
measuring devices: trained in instrument safety 
– familiar with the individual protection concepts of the devices
– for Ex-certified measuring devices: also trained in explosion protection

• The spare parts set and Installation Instructions are used to replace a faulty unit 

with a functioning unit of the same type.
Use genuine parts from Hauser only. 

• Using the material number and identification information, check whether the 

spare part matches the identification label on the measuring device.

• If you have any questions, contact your E+H service organization.
• Modifications of the measuring device are not permitted.
• Follow all measuring device documentations.
• In the case of Ex-certified measuring devices: Comply with the operating and  

installation instructions and operating conditions in the supplementary 
Ex-documentation. 

• Hot surfaces! Risk of injury! 

Before commencing work, allow the system and measuring device to cool 
down to a touchable temperature.

• In the case of measuring devices in custody transfer applications, the custody 

transfer status no longer applies once the lead seal has been removed.

• In the case of measuring devices in safety-related applications in accordance 

with EN 61508 or EN 61511, reconfigure and test following any repair. 
Document the repair procedure in accordance with the applicable standard(s).

• In the case of Ex-certified measuring devices: Only open in a de-energized 

state (once a delay of 10 minutes has elapsed after switching off the power 
supply) or in environments which do not have a potentially explosive 
atmosphere.

• After removing the electronics cover, there is a risk of electric shock as shock 

protection is removed!
Switch off the measuring device before removing internal covers.

• Service connector: 

– do not connect in potentially explosive atmospheres.
– only connect to Hauser service devices.

• Only open housing for a brief period. Avoid the penetration of foreign bodies,  

moisture or contaminants.

• To regain the (explosion) protection of the measuring device:

– seal the openings with genuine covers
– seal cable entries with approved components suitable for the (explosion) 

protection.

• Replace defective seals/gaskets by genuine parts from Hauser only.
• Threads (e.g. of the cover for the electronics and connection compartments) 

must be lubricated. Use an acid-free, non-hardening grease if an abrasion 
resistant dry lubricant is non-existent. 

• If threads are damaged or defective, the measuring device must be repaired.
• If spacing is reduced or the dielectric strength of the measuring device cannot 

be guaranteed during repair work, perform a test on completion of the work 
(e.g. high-voltage test in accordance with the manufacturer's instructions). 

• Observe the instructions for transporting and returning the device outlined in 

the Operating Instructions.

2 Replacing cover with O-Ring and claw 

3 Re-assembly

Re-assembly is carried out in reverse order, unless otherwise instructed. 
The following must be noted:

a

A0012820

b

A0012822

a

A0011687

3 mm