Installation
Proline t-mass 65 PROFIBUS DP/PA
18
Hauser
4.2.6
Heating
Some applications require suitable measures to avoid heat loss (condensation). Heating can
be electric, e. g. with heated elements, or by means of hot water, steam pipes or insulation.
"
Caution!
Risk of electronics overheating! Consequently, make sure that the adapter between sensor
and transmitter and the connection housing of the remote version always remain free of
insulating material.
4.2.7
Thermal insulation
When the gas is very damp or saturated with water (e. g. bio gas), the piping and flowmeter
body should be insulated to prevent water droplets condensing on the measuring sensor.
A0005122
Fig. 8:
Maximum thermal insulation for t-mass 65F and 65I
a
Maximum insulation height for the flanged sensor
b
Maximum insulation height for the insertion sensor
4.2.8
Vibrations
"
Caution!
Excessive vibration can result in mechanical damage to the measuring device and its mount-
ing.
Observe the vibration spezification in the technical data section
a
a
b
b