Proline Prowirl 73
Installation
Hauser
9
2.2.5 Inlet and outlet runs
The following inlet and outlet runs must be taken into account. The longest inlet run shown
must be observed if two or more flow disturbances are present.
A0001867
Perforated plate flow conditioner
A specially designed perforated plate flow conditioner, available from Hauser, can be
installed if it is not possible to observe the inlet runs required. The flow conditioner is fitted
between two piping flanges and centered with the mounting bolts. Generally, this reduces the
inlet run required to 10 × DN with complete accuracy.
A0001887
A = Inlet run
B = Outlet run
1 = Reduction
2 = Extension
3 = 90° elbow or T-section
4 = 2 × 90° elbow, three-dimensional
5 = 2 × 90° elbow
6 = Control valve
15 x DN
5 x DN
A
1
3
5
2
4
6
A
A
A
A
A
B
B
B
B
B
B
17 x DN + 8 x h
5 x DN
20 x DN
5 x DN
40 x DN
5 x DN
25 x DN
5 x DN
50 x DN
5 x DN
h
8 x DN
2 x DN
5 x DN