Proline Prosonic Flow P 500
Hauser
27
Order code for "Accessory mounted", option NB "Adapter RJ45 M12 (service interface)"
Order code
Cable entry/coupling → 28
"Accessory mounted"
Cable entry
2
Cable entry
3
NB
Plug M12 × 1
–
Pin assignment, device plug
Service interface
Order code for "Accessories mounted", option
NB
: Adapter RJ45 M12 (service interface)
3
2
4
1
A0032047
Pin
Assignment
1
+
Tx
2
+
Rx
3
-
Tx
4
-
Rx
Coding
Plug/socket
D
Socket
Recommended plug:
• Binder, series 763, part no. 99 3729 810 04
• Phoenix, part no. 1543223 SACC-M12MSD-4Q
Supply voltage
Order code for
"Power supply"
Terminal voltage
Frequency range
Option
D
DC 24 V
±20%
–
Option
E
AC 100 to 240 V –15 to +10%
50/60 Hz, ±4 Hz
Option
I
DC 24 V
±20%
–
AC 100 to 240 V –15 to +10%
• 50/60 Hz
• 50/60 Hz, ±4 Hz
Power consumption
Transmitter
Max. 10 W (active power)
switch-on current
Max. 36 A (<5 ms) as per NAMUR Recommendation NE 21
Current consumption
Transmitter
• Max. 400 mA (24 V)
• Max. 200 mA (110 V, 50/60 Hz; 230 V, 50/60 Hz)
Power supply failure
• Totalizers stop at the last value measured.
• Depending on the device version, the configuration is retained in the device memoryor in the
pluggable data memory (HistoROM DAT).
• Error messages (incl. total operated hours) are stored.
Electrical connection
Connection of the connecting cable: Proline 500
The connecting cable is connected via terminals.