Proline Prosonic Flow 91
Installation
Hauser
29
a0005875
Fig. 29:
Connecting nameplate for sensor connecting cables
3.10.3
Cable specifications
Sensor cable
• Use the ready-to-use cables supplied by Hauser with each sensor pair.
• The following cable lengths are available:
– 5 m, 10 m, 15 m , 30 m, 60 m
– 16 ft, 33 ft, 49 ft, 98 ft, 197 ft
• Cable material:
PVC (DN 50 to 4000/ 2" to 160")
TPE-V (DN 15 to 65 / ½" to 2½")
• Operating temperature: –20 to +70 °C (–4 to +158 °F)
Operation in zones of severe electrical interference:
The measuring device complies with the general safety requirements in accordance with
EN 61010-1, the EMC requirements of IEC/EN 61326 "Emission to class A requirements", and
NAMUR Recommendation NE 21.
UP
DOWN
Field-
Check
T
-DA
T