Measurement Report
If required, the device can be supplied with a measurement report. To certify the performance of
the device, a measurement is performed under reference conditions. Here, the sensors are mounted
on a pipe with a nominal diameter of DN 15 (½"), DN 25 (1"), DN 40 (1½"), DN 50 (2") or
DN 100 (4") respectively.
The measurement report guarantees the following error limits of the device
[at a flow velocity > 0.3 m/s (1 ft/s) and a Reynolds number > 10000]:
Repeatability
Max. ± 0.3% for flow velocities > 0.3 m/s (0.98 ft/s)
9.1.7
Operating conditions: Installation
Installation instructions
Any orientation (vertical, horizontal)
Restrictions and additional installation instructions
Inlet and outlet runs
Clamp-on version
Length of connecting cable
Shielded cables are offered in the following lengths:
5 m, 10 m, 15 m, 30 m
60 m only available for sensors DN50...4000
15 feet, 30 feet, 45 feet, 90 feet
180 feet only available for sensors DN 2" to 160"
Route the cable well clear of electrical machines and switching elements.
Sensor
Nominal diameter
Guaranteed error limits of the device
Prosonic W
DN 15 (½"), DN 25 (1"), DN 40 (1½"), DN 50 (2")
±0.5 % o.r. ± 5 mm/s
Prosonic W
DN 100 (4")
±0.5 % o.r. ± 7.5 mm/s
o.r. = of reading